90088262 MCS130B
CORNERSTONE 130B MONOCHROMATORS
54
Command:
GRATing
Purpose:
Store the grating setting and rotate the grating stage to the desired
grating (if it’s different from the current grating). The stage will rotate to
the location required for the new grating to pass the currently-selected
wavelength to the output.
Parameter:
A decimal number from 1 to 2.
Example:
grating 2
: if the current grating is not 2, rotate the grating stage to
grating 2.
Default:
1
Command:
GRATing?
Purpose:
Query the current grating setting.
Example:
grating?
Example:
grat?
E.g. response
1
: the currently-selected grating is number 1.
Command:
GRAT1FACTOR
Purpose:
Set the calibration multiplier for grating 1.
Parameter:
A floating-point number.
Example:
Grat1factor 1.1
: set the cal factor for grating 1 to 1.1.
Example:
grat1factor 1
: set the cal factor for grating 1 to 1.0.
Default:
1.0.
Command:
GRAT1FACTOR?
Purpose:
Query the calibration multiplier for grating 1.
Example:
grat1factor?
E.g. response
0.99
: the cal factor is 0.99.
Command:
GRAT1LABEL
Purpose:
Set a label for grating 1.
Parameter:
An unquoted string up to 8 characters in length.
Example:
grat1label blz800
: set the label for grating 1 to “blz800”.
Default:
“ ” (spaces).
Command:
GRAT1LABEL?
Purpose:
Query grating 1’s label.
Example:
grat1label?
E.g. response
Hitone
: the label has been set to the string “Hitone”.
Содержание Cornerstone 130B
Страница 1: ...90088262MCS130B Rev B s 1 8m Monochromator Family User s Manual Cornerstone 130B...
Страница 32: ...90088262 MCS130B CORNERSTONE 130B MONOCHROMATORS 32 Figure 23 Model 74010 Filter Wheel...
Страница 42: ...90088262 MCS130B CORNERSTONE 130B MONOCHROMATORS 42 14 13BDIMENSIONS Figure 30 Cornerstone 130B Dimensions...
Страница 44: ...90088262 MCS130B CORNERSTONE 130B MONOCHROMATORS 44 Figure 33 Model 74006 Mounting Plate...