90088262 MCS130B
CORNERSTONE 130B MONOCHROMATORS
38
11.5
49B
GRATING INSTALLATION
Each grating comes mounted and aligned on a grating mount. For most applications, there is no
need to adjust this grating holder. Refer to the general warnings and precautions section of this
manual regarding the care and handling of diffraction gratings. Note that only one grating should
be replaced at a time.
Procedure:
1. Using the Cornerstone Hand Controller or utility software, close the shutter.
2. Command the Cornerstone to go to the grating that is to be replaced.
3. Go to wavelength 0. The grating facing the collimating mirror is the grating that was selected.
4. Turn off the power switch on the monochromator and unplug the instrument from the
electrical mains.
5. Remove all screws holding the monochromator cover in place, then remove the cover.
6. Two long screws, one on each side of the grating, must be removed. Used a screwdriver t o
loosen these screws.
7. Grasp the sides of the grating mount with one hand and firmly pull the grating up and out of
the instrument. Pull in a direction parallel to the grating face
– not directly upwards . During
this process, ensure the other grating does not move. It should be held in place using a free
hand by grasping or pressing down on the grating holder.
8. Carefully uncover the new grating to be installed and set aside.
9. Use the cover from the new grating to protect the grating that was just removed.
10. Install the new grating onto the grating platform and ensure the new grating is parallel t o t he
other grating. The grating should fit into a pin located on the platform.
11. Push the new grating up against the large central pin in the middle of the platform.
12. Re-install the long screws that held the previous grating in place. Use only firm finger
pressure to tighten them. Do not tighten one completely while the other screw is loose. It is
best to alternate tightening between the two screws. Ensure the screws line up parallel wit h
the grating face.
13. Replace the cover on the monochromator and secure using all screws provided. Failure t o
use all hardware will result in increased stray light.
14. Connect the monochromator to the electrical mains and switch on the power.
15. Follow the procedures outlined in this manual to determine the grating factor and any offset.
16. Use the Mono utility software to update the calibration parameters as needed.
17. It is suggested to verify the calibration factors and offsets of the other gratings that were not
replaced, as they may have accidentally been moved during this process.
Содержание Cornerstone 130B
Страница 1: ...90088262MCS130B Rev B s 1 8m Monochromator Family User s Manual Cornerstone 130B...
Страница 32: ...90088262 MCS130B CORNERSTONE 130B MONOCHROMATORS 32 Figure 23 Model 74010 Filter Wheel...
Страница 42: ...90088262 MCS130B CORNERSTONE 130B MONOCHROMATORS 42 14 13BDIMENSIONS Figure 30 Cornerstone 130B Dimensions...
Страница 44: ...90088262 MCS130B CORNERSTONE 130B MONOCHROMATORS 44 Figure 33 Model 74006 Mounting Plate...