Annex 2. Electromagnetic Emissions and Immunity
37
Guidance and manufa
cturer’s declaration – noise immunity
The system is intended for use in the electromagnetic environment specified
below. The customer or the user of the system should assure that it is used in
such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment - guidance
Portable
and
mobile
RF
communications
equipment
should be used no closer to
any part of the system,
including cables, than the
recommended
separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
The recommended separation
distance:
Conducted RF
IEC 61000-4-
6:2013
3 V in 0,15
MHz-80 MHz
band; 6 V in
ISM band
between 0,15
MHz and 80
MHz; 80% AM
at 1 kHz
1)
3 V
3)
P
d
17
.
1
Radiated RF
IEC 61000-4-
3:2006
10 V/m
80 MHz
– 2,7
GHz
80 % AM at 1
kHz
3 V/m in 80
MHz-2.7 GHz
band, 80% AM
at 1 kHz
10 V/m
P
d
17
.
1
(80 MHz to 800 MHz);
P
d
33
.
2
(800 MHz to 2.7 GHz),
Field strengths from fixed RF
transmitters, as determined by
an
electromagnetic
site
survey
1)
, should be less than
the compliance level in each
frequency range
2)
. Interference
may occur in the vicinity of
equipment
marked
with
the following symbol: