Assembly and Installation
17
It is recommended to place the electronic unit on the maxi-
mum possible distance from power cables, switchboards,
and different powerful electrical devices which can emit elec-
tromagnetic fields of mains frequency.
Requirements to mains:
Do not use electric mains where the neutral conductor and
protective ground are combined. It is strongly prohibited.
Before the system is installed, the quality of standard tripolar
sockets and the integrity of protective ground circuit must be
checked by electrician.
The recommended distance from the electronic unit to the
nearest electric mains is not less than 3 meters.
2.3. Unpacking and Check of Delivery
Set
In case the box with the system was under conditions of the
excessive moisture or low temperature which differs vastly from
the working conditions, place the system in the room with nor-
mal conditions and leave it there for 24 hours.
Unpack the box and take the system components out. The
delivery set should correspond to the packing report.
The computer equipment packed in the separate boxes should
be opened according to user and technical manuals for these
products.
Check the system components and make sure that there is no
external damage.