Neuro-ERG/V (Technical Manual)
30
gen peroxide solution). The use of organic solvents and aromatic oils must be avoid-
ed. Never submerge the device or the cables in disinfectant or other liquids.
After testing the electrodes should be removed from the examined animal immediate-
ly. The disposable electrodes should be discarded after their use. The reusable elec-
trodes should be disinfected in accordance with the accepted standards of veterinary
practice.
5.
Current Repair
5.1. General Requirements
The repair of the system requires special training of technical staff, special equipment
and service software. The manufacturer and its representatives have everything nec-
essary for this. The repair connected with the opening of electronic units is prohibited.
The repair of computer equipment can be performed by companies specialized in
computer equipment servicing.
The current repair of the system includes the repair of some component parts and ca-
bles. If the component parts are connected to the system, the repair is prohibited.
When performing the repair, all units must be switched off.
5.2. Cables and Adapters
The cables are examined externally, and the circuit is checked for short circuit or
break. In case the cable is broken or short-circuited, replace it or cut it if the cable
length is sufficient.
5.3. Computer Interface Cable (USB Cable)
The computer interface cable (Fig. 15) is examined externally, and the circuit is
checked for short circuit or break. In case the cable is broken, replace it or cut it if the
cable length is sufficient. Please, pay attention for the cable marking. The marking of
wire
size
on
the
cable
should
be
either
28AWG/2C+24AWG/2C,
or
28AWG/2C+22AWG/2C, or 28AWG/2C+20AWG/2C.