![Neurosoft Neuro-ERG/V Скачать руководство пользователя страница 21](http://html1.mh-extra.com/html/neurosoft/neuro-erg-v/neuro-erg-v_technical-manual_1676386021.webp)
Assembly and Installation
21
Fig. 12. Placement of equipment connected to notebook PC.
As a power unit, the notebook PC power unit corresponding to IEC 60601-1:2012
(AAMI/ANSI ES 60601-1:2005/(R2012) and A1:2012 and A2:2010/(R)2012, EN
60601-1:2006/A1:2013) is used. As for isolation transformer, the isolation transformer
ТМ-630М or its analogue corresponding to IEC 60601-1:2012 (AAMI/ANSI ES 60601-
1:2005/(R2012) and A1:2012 and A2:2010/(R)2012, EN 60601-1:2006/A1:2013) can
be used.
It is allowed to connect the system components to PC using the external USB hub
КМ-7 or its analogue that correspond to IEC 60601-1:2012 (AAMI/ANSI ES 60601-
1:2005/(R2012) and A1:2012 and A2:2010/(R)2012, EN 60601-1:2006/A1:2013).
The USB hub is connected to the network in accordance with Fig. 12.