Neuro-ERG/V (Technical Manual)
24
3.
Proper use
3.1. Safety Measures
To provide safety measures and exclude the possibility of electric trauma of
medical staff or examined animal, it is PROHIBITED:
to use the system which mounting and setting was done incorrectly without follow-
ing the manual instructions;
to connect the system and surgical HF equipment to the examined animal (it can
lead to the damage of the system or cause flash-burns in the places of electrode
placement);
to connect any devices, not included in the system delivery set, to the electrode
jacks;
to eliminate faults by opening of the system components included in the delivery
set;
to perform tests when the electronic unit, computer or other devices of the system
are opened;
to connect electrodes placed on the examined animal to
protective ground
or
other conducting surfaces.
3.2. Getting Ready
Operating Limitations:
Ambient temperature is from +10 to +35
С.
Relative humidity is from 30 to 85% (non-condensing).
Atmospheric pressure is from 70 to 106 kPa.
Before the power supply is switched on, make sure that the electronic unit and com-
puter equipment cases have no apparent mechanical failures which can represent
danger.
Power Supply Switch on and System Test
The power supply of the system is switched on by pressing the
Power
button on the
PC. The electronic unit has no power supply and it is connected to the PC all the time.
The power supply is switched on when the operating system is loaded and the
Neuro-
MEP.NET
program is started.