Neurosoft Neuro-ERG/V Скачать руководство пользователя страница 27

Proper use 

27 

Table 3. Continued

 

Trouble Symptom 

Cause 

Way of Removal 

High 

electromagnetic 

interference  in  the  recorded 
signal  (frequency 

–  50  Hz  or 

divisible  to  it 

–  100  Hz,  150 

Hz, etc.). 
 

The  electrodes  not  used  at 
the  moment  are  connected  to 
the 

amplifier 

(i.e. 

the 

electrodes  are  not  applied  on 
the  examined  animal  but 
“hang  in  the  air”  and  cause 
interference). 

Disconnect the electrodes  not 
used  at  the  moment  from  the 
amplifier. 

Improper 

placement 

of 

electrodes. 

Start 

the 

impedance 

measurement  and  obtain  the 
accepted  value.  In  case  of 
electrode 

failure, 

replace 

them  (for  example,  in  the 
event of a break). 

Bad grounding of the system.  Ground the system properly. 

The  presence  of  powerful 
noise 

sources 

(X-ray 

equipment,  physiotherapeutic 
room,  powerful  refrigerating 
systems, 

electric 

motors, 

electric welding, etc.). 

Try  to  switch  off  or  move  off 
these powerful devices on the 
bigger 

distance. 

If 

it 

is 

impossible, try to obtain better 
parameters  of  the  system 
grounding. 

There  is  no  stimulus  during 
the 

acquisition 

with 

stimulation. 

The stimulation device is not 
connected (or connected 
unsafe) to the stimulator unit. 

Make sure that the stimulation 
device is connected to the 
stimulator unit safely. 

The connection between 
stimulator and computer is 
broken. 

Execute the 

Setup|Hardware 

reset

 menu command and try 

the start again. If it fails, 
restart the computer. 

Содержание Neuro-ERG/V

Страница 1: ...Technical Manual Neuro ERG V Veterinary Digital ERG System TM006 04 001 000 14 06 2019...

Страница 2: ...Neurosoft 2020 5 Voronin str Ivanovo 153032 Russia P O Box 10 Ivanovo 153000 Russia Phone 7 4932 24 04 34 7 4932 95 99 99 Fax 7 4932 24 04 35 E mail info neurosoft com Internet www neurosoft com...

Страница 3: ...19 2 3 Unpacking and Check of Delivery Set 22 2 4 System Assembly and Connection 22 3 Proper use 24 3 1 Safety Measures 24 3 2 Getting Ready 24 3 3 Troubleshooting 25 3 4 Getting Started 28 3 5 Action...

Страница 4: ...ns to the following address P O Box 10 Ivanovo 153000 Russia or by e mail help neurosoft com You can find additional information about Neurosoft products on our website www neurosoft com or ask questi...

Страница 5: ...nal functions with altered incomplete missing pupil response to red light during normal response to blue light and also if it is impossible to evaluate the response to light atrophy of iris miosis dia...

Страница 6: ...inflammatory and or allergic disease of cornea conjunctiva eyelid or traumatic injury of eyeball contraindications for sedation Possible Side Effects Usually the side effects are not observed However...

Страница 7: ...leakage current not more than 0 1 A Bandpass flatness in the band from 0 02 up to 0 05 Hz and from 5 up to 10 kHz from 0 05 Hz up to 5 kHz from 30 up to 5 from 10 up to 5 High pass filter 0 02 0 05 0...

Страница 8: ...right two sided stimulation when using LED goggles available General Parameters and Characteristics Interface USB Supply voltage electronic unit desktop PC based system notebook PC based system 5 V DC...

Страница 9: ...ortable and mobile RF communication equipment can affect the system operation The use of equipment not listed in Table 5 of this technical manual may result in in creased emission and system decreased...

Страница 10: ...nd are transferred to the microcontroller MC via 4000 V galvanic isola tion The microcontroller provides the connection with the computer via USB and the transfer of the digitized data to the computer...

Страница 11: ...ic stimulator is LED goggles with a set of super power LEDs for the separate stimulation of the left and right eye All the units are attached to the computer via USB hub The system operates under cont...

Страница 12: ...amplifier unit there are the touch proof and DIN connectors to attach the electrodes and operation indicator Fig 2 The channel numbers are marked with Arabic figures 1 2 3 and 4 The operation indicat...

Страница 13: ...mulator unit are shown in Fig 4 and Fig 5 On the front panel of the auditory visual stimulator unit there are the connectors for visual stimulator LED goggles or mini ganzfeld multi colored penlights...

Страница 14: ...lso its operation is described in the corresponding technical manual 1 4 Synchronization with Stimulators of Third party Manufacturers The trigger input socket is used for synchronization of the ampli...

Страница 15: ...pulse appears at the connector which fall corresponds to the beginning of the stimulus The pulse duration at the trigger output is 2 5 s 1 5 Synchronization with Neuro MEP NET Software The external s...

Страница 16: ...i from the external stimulator are ignored 6 Start stimulation from the external stimulator In other aspects the signal acquisition does not differ from the one described in the corresponding chapters...

Страница 17: ...es relating to electromagnetic compatibility applied parts of BF type according to IEC 60601 1 2012 AAMI ANSI ES 60601 1 2005 R2012 and A1 2012 and A2 2010 R 2012 EN 60601 1 2006 A1 2013 This symbol i...

Страница 18: ...trix is a two dimensional matrix barcode consisting of black and white cells or modules of different brightness arranged in either a square or rectangular pattern The DataMatrix barcode is described i...

Страница 19: ...s unstable operation It is recommended to place the electronic unit on the maximum possible distance from power cables switchboards and different powerful electrical devices which can emit electromagn...

Страница 20: ...everal tens of amperes compensating current leakage through the system connecting cables that leads to the equipment break down Requirements for PC connection to local area network LAN To avoid the el...

Страница 21: ...ansformer the isolation transformer 630 or its analogue corresponding to IEC 60601 1 2012 AAMI ANSI ES 60601 1 2005 R2012 and A1 2012 and A2 2010 R 2012 EN 60601 1 2006 A1 2013 can be used It is allow...

Страница 22: ...re the first connection of the system to PC Read the corresponding section of the user manual before starting to work If the distributive is missing or the software update is required address to your...

Страница 23: ...lace the holder as near as possible to the place of exam carrying out and fix the am plifier unit on it Connect all the necessary equipment The electrical units can be con nected to PC when the power...

Страница 24: ...erform tests when the electronic unit computer or other devices of the system are opened to connect electrodes placed on the examined animal to protective ground or other conducting surfaces 3 2 Getti...

Страница 25: ...er The program message Connection error Neuro MEP is already used by another program The other program using the system is opened for example the second copy of Neuro MEP NET program Close the running...

Страница 26: ...re connected to the different USB controllers The amplifier and stimulator are connected to the USB connectors of the PC referring to different USB controllers Usually such jacks are located far from...

Страница 27: ...e in the event of a break Bad grounding of the system Ground the system properly The presence of powerful noise sources X ray equipment physiotherapeutic room powerful refrigerating systems electric m...

Страница 28: ...e electrodes are applied check the impedance to control the quality of their placement First of all the impedance measurement mode should be switched on see the user manual The signal acquisition and...

Страница 29: ...from the surface of the cases using wet fabric and also disinfection according to section 4 3 4 3 Disinfection Before cleaning the electronic unit switch it off As you clean visually inspect the unit...

Страница 30: ...bited The repair of computer equipment can be performed by companies specialized in computer equipment servicing The current repair of the system includes the repair of some component parts and ca ble...

Страница 31: ...ector side is performed according to the schematic shown in Fig 16 The check is done using the device designed to test the diodes Disassemble the housing of the cable connector and inspect the montage...

Страница 32: ...r USB cable re pair section 5 3 of this manual The detection and elimination of LEDs failure can be done only by the manufacturer Fig 17 Electrical schematic of LED penlights 5 6 Mini Ganzfeld Stimula...

Страница 33: ...llion content data Name Silver content g Corneal hook electrode NS006106 004 0 21 8 Delivery Set and Package Data The Neuro ERG V veterinary digital ERG system is collected and packed The system is re...

Страница 34: ...sportation mounting and technical servicing are not observed the warranty period is expired user breaks the seal without permission of the manufacturer 9 4 The manufacturer is obliged to repair the sy...

Страница 35: ...ble please indicate the reasons and cir cumstances preceding the fault detection in addition it is recommended to add the test report the test data photos and other materials allowing solving the prob...

Страница 36: ...equency 2402 2480 MHz 1 USB cable A B NS007103 005 01 1 Technical manual for dedicated keyboard TM035 02 001 000 1 Package NS035901 001 1 Electrode holder NS016221 023 1 Footswitch NS028353 005 1 USB...

Страница 37: ...n transformer TM 630 1 NS036201 004 1 Cable IEC C13 IEC C14 220 V 10 A L 1 8 m 3G 0 75 sq mm 1 Power cable IEC C14 Schuko type F socket 220 230 V 10 A L not more than 1 m 3G 0 75 sq mm 1 Technical man...

Страница 38: ...nt guidance RF emissions CISPR 11 Group 1 The system uses RF energy only for its internal function Therefore its RF emissions are very low and not likely to cause any interference in nearby electronic...

Страница 39: ...ode 0 5kV 1 kV 2 Mains power quality should be that of typical commercial or hospital environment 2 kV common mode 0 5kV 1 kV 2kV 2 Voltage dips IEC 61000 4 11 2004 0 UT during 0 5 period at 0 45 90 1...

Страница 40: ...tion equipment IEC 61000 4 3 2006 Band MHz Immunity level V m The use of portable and mobile RF communications equipment can adversely affect the recording do not use an operating cellular phone withi...

Страница 41: ...equency of transmitter d m P d 17 1 in the band from 150 kHz up to 80 MHz P d 17 1 in the band from 80 up to 800 MHz P d 33 2 in the band from 800 MHz up to 2 7 GHz 0 01 0 117 0 117 0 233 0 1 0 369 0...

Отзывы: