Diagnostic selections:
•
ROM (
F1
): Performs a ROM checksum to verify the integrity of
the processor's memory. A pass/fail message will display. This
function is also performed on power-up.
•
EP (
F2
): Performs a non-volatile memory (EEPROM) checksum
to verify the integrity of the machine calibration constants. A
pass/fail message will display. This function is also performed on
power-up.
Notes on proper operation
•
When a change in temperature selection is made (either Blood
Temp. or Inj. Temp.), the simulated resistance outputs
experience a small amount of fluctuation. For best results, allow
10 seconds for the baseline to stabilize before starting a
simulation.
•
The battery voltage is continually monitored during normal
operation. If a low battery condition is detected, a “LOW
BATTERY” message will appear and all functions will be
disabled.