General Overview 

 

 

Section 1

 

 

1.1 Introduction 

 

The 

MICROSIM COS

 is a compact cardiac out put simulator, designed 

to test and verify the functionality of most cardiac output monitors. It simulates 
an ideal cardiac output waveform for various preset values for selected Blood 
temperature and injectate temperatures.  

 

The 

MICROSIM COS

 utilizes a proprietary complex software algorithm 

to generate a realistic cardiac output waveform. It is recommended and used 
by many Medical equipment manufacturers to verify the performance of their 
cardiac out put monitors.  

 

It is compact, lightweight, and rugged.  The 

MICROSIM COS

 is 

operated via easy to use menus displayed on its LCD display.  It is battery 
operated.

 

 

1.2 Specifications 

 

Catheter Size:

 7F 

Injectate Volume:

 10cc 

Computational Constant:

 0.542 

 

Cardiac output Selections

: 3, 5, 6, 7 LpM (letter/minute) 

Accuracy 2% +/- .1 LpM 

 

Blood Temperature

: 37

°

C and 38

°

Accuracy 2% +/- 0.1

°

 

Injectate Temperature

: 5

°

C and 20

°

Accuracy 2% +/- 0.1

°

 

Electrical: 

  

Power: 9 Volt Alkaline Battery  
Battery life: 100 hours Continues operation 

 

Mechanical: 

  

Size: 4.75x2.75x1.25” 
Unit Weight: .5Lbs with Battery 
Shipping Weight: 3 Lbs. 

 

1.3  

Accessories (Standard)

 

 

PN: 11101 

Baxter Edwards Compatible Interface Cable 

PN: 11104 

Carrying case 

 

Various Interface cables are available. Contact Netech.

 

Содержание MicroSim COS

Страница 1: ...COS Cardiac Output Simulator The one and only that simulates a true waveform Operating Manual Netech Corporation 110 Toledo Street Farmingdale NY 11735 Phone 631 531 0100 Fax 631 531 0101 http www Ne...

Страница 2: ...ue the MICROSIM COS at any time and change its specifications price or design without notice and without incurring any obligation Netech guarantees violability of service parts for 5 years after the m...

Страница 3: ...ble for partial refund and or credit Non serialized parts and accessory items e g cables carrying cases auxiliary modules etc are not eligible for return or refund Only products returned within 90 day...

Страница 4: ...er service dept before a product is returned for refund or credit The RMA number should be clearly marked on the package along with the a statement indicating the reason for return Repair and Calibrat...

Страница 5: ...d Line Conducted Emissions Verification of compliance was conducted to the limits and methods of the following CISPR 16 1 1993 and CISPR 16 2 1996 Immunity The system has been type tested and found to...

Страница 6: ...EEC Low Voltage The system has been type tested and found to meet the requirements of EC Directive 73 23 EEC for Low Voltage Verification of compliance was conducted to the limits and methods of the...

Страница 7: ...Interface 5 EC Directive 6 Table of Contents 7 1 General Overview 1 1 Introduction 8 1 2 Specifications 8 1 3 Accessories 2 Operating Instructions 2 1 Getting Started 9 2 2 Instrument Familiarity 9 2...

Страница 8: ...out put monitors It is compact lightweight and rugged The MICROSIM COS is operated via easy to use menus displayed on its LCD display It is battery operated 1 2 Specifications Catheter Size 7F Inject...

Страница 9: ...e Netech warranty statement is listed in the warranty section of this manual When shipping an instrument to Netech for repair or calibration make sure that the instrument is properly packed A complete...

Страница 10: ...elections The display consists of a 16x2 character LCD display The menu structure is as follows Initialization Menu On power up the initialization screen will appear N E T E C H C O S 2 0 0 0 Internal...

Страница 11: ...e c t i o n C O I T B T C f g F1 F2 F3 F4 Active keys in the Selection Menu CO F1 Enters the CO rate selection menu IT F2 Enters the Injectate Temperature selection menu BT F3 Enters the Blood Temper...

Страница 12: ...resistance of an ideal Catheter thermistor The two selections available are 37 C 98 6 F and 38 C 100 4 F The current setting is displayed Selecting a new value returns you to the Main Screen Press Es...

Страница 13: ...essage will display This function is also performed on power up Notes on proper operation When a change in temperature selection is made either Blood Temp or Inj Temp the simulated resistance outputs...

Страница 14: ...ired blood temperature injectate temperature and CO rate settings as outlined in section 2 3 5 Allow 10 seconds for the baseline to stabilize 6 Press the Start button on the monitor under test 7 Press...

Страница 15: ...umper short This is a calibration procedure error It is meant as a reminder to disable calibration before leaving a calibration lab If this error should appear during normal operation an internal faul...

Страница 16: ...16 Appendix Section 4...

Отзывы: