 10

 

2.3 Controls and Menu Structure 

 

Controls 

 
The control consists of four function keys 

F1

 – 

F4

, a left arrow 

(

<

), a right arrow (

>

), 

Esc

 key and power On/Off. 

 
Each function key is active based on the context of each menu. 
Function keys are assigned to selections appearing along the 
bottom line of the display. When there are more selections than 
can fit on one screen, left and right arrows will be displayed, and 
the arrow keys become active to permit scrolling of selections. 
For most menus, the 

Esc

 key returns you to the Main Screen 

without making any selections. The display consists of a 16x2 
character LCD display. The menu structure is as follows. 

 

Initialization Menu 

 

On power-up, the initialization screen will appear: 

 

          N  E  T  E  C  H           
        C O S    2  0  0  0         

 

Internal tests are performed during power on initialization, If all 
initialization  tests pass, the unit will move on to the Simulator 
Main menu as below. 

 

Simulator Main Screen 

 

B  T  : 3  7  °  C      I  T  :  5  °  C 
C O  : 5  L  p m     S  e  l  S  t r 

  

 

F3 

 

F4 

 

 

Menu Abbreviations

:  

BT: Blood Temperature 
IT: Injectate Temperature 
CO: Cardiac Output 
Sel: Selection menu 
Str: Start waveform 

 
Active selection keys in the Main Screen are F3 and F4: 

 

 (

F3

): Selects other CO values. 

 The default value is set at 5  LpM. 

 (

F4

): Starts the CO waveform. 

 
Press (

F4

) to start simulation. A progress bar will appear at the 

Содержание MicroSim COS

Страница 1: ...COS Cardiac Output Simulator The one and only that simulates a true waveform Operating Manual Netech Corporation 110 Toledo Street Farmingdale NY 11735 Phone 631 531 0100 Fax 631 531 0101 http www Ne...

Страница 2: ...ue the MICROSIM COS at any time and change its specifications price or design without notice and without incurring any obligation Netech guarantees violability of service parts for 5 years after the m...

Страница 3: ...ble for partial refund and or credit Non serialized parts and accessory items e g cables carrying cases auxiliary modules etc are not eligible for return or refund Only products returned within 90 day...

Страница 4: ...er service dept before a product is returned for refund or credit The RMA number should be clearly marked on the package along with the a statement indicating the reason for return Repair and Calibrat...

Страница 5: ...d Line Conducted Emissions Verification of compliance was conducted to the limits and methods of the following CISPR 16 1 1993 and CISPR 16 2 1996 Immunity The system has been type tested and found to...

Страница 6: ...EEC Low Voltage The system has been type tested and found to meet the requirements of EC Directive 73 23 EEC for Low Voltage Verification of compliance was conducted to the limits and methods of the...

Страница 7: ...Interface 5 EC Directive 6 Table of Contents 7 1 General Overview 1 1 Introduction 8 1 2 Specifications 8 1 3 Accessories 2 Operating Instructions 2 1 Getting Started 9 2 2 Instrument Familiarity 9 2...

Страница 8: ...out put monitors It is compact lightweight and rugged The MICROSIM COS is operated via easy to use menus displayed on its LCD display It is battery operated 1 2 Specifications Catheter Size 7F Inject...

Страница 9: ...e Netech warranty statement is listed in the warranty section of this manual When shipping an instrument to Netech for repair or calibration make sure that the instrument is properly packed A complete...

Страница 10: ...elections The display consists of a 16x2 character LCD display The menu structure is as follows Initialization Menu On power up the initialization screen will appear N E T E C H C O S 2 0 0 0 Internal...

Страница 11: ...e c t i o n C O I T B T C f g F1 F2 F3 F4 Active keys in the Selection Menu CO F1 Enters the CO rate selection menu IT F2 Enters the Injectate Temperature selection menu BT F3 Enters the Blood Temper...

Страница 12: ...resistance of an ideal Catheter thermistor The two selections available are 37 C 98 6 F and 38 C 100 4 F The current setting is displayed Selecting a new value returns you to the Main Screen Press Es...

Страница 13: ...essage will display This function is also performed on power up Notes on proper operation When a change in temperature selection is made either Blood Temp or Inj Temp the simulated resistance outputs...

Страница 14: ...ired blood temperature injectate temperature and CO rate settings as outlined in section 2 3 5 Allow 10 seconds for the baseline to stabilize 6 Press the Start button on the monitor under test 7 Press...

Страница 15: ...umper short This is a calibration procedure error It is meant as a reminder to disable calibration before leaving a calibration lab If this error should appear during normal operation an internal faul...

Страница 16: ...16 Appendix Section 4...