11
lower right as simulation starts. It provides a visual indication of
the percentage of the elapsed waveform. During a simulation,
the COS will not respond to any keys pressed. To show the
simulation has completed, the progress bar will become a series
of dashes as shown:
B T : 3 7 ° C I T : 5 ° C
C O : 5 L p m - - - - - - -
Pressing
F3
,
F4
, or
Esc
will reset the simulator and return to the
Main Screen.
Selection
Menu
Press (
F3
)
‘Sel’
from the Main Screen to enter the Selection
Menu:
S e l e c t i o n
C O I T B T C f g
F1 F2
F3 F4
Active keys in the Selection Menu:
•
CO (
F1
): Enters the CO rate selection menu.
•
IT
(
F2
): Enters the Injectate Temperature selection menu.
•
BT (
F3
): Enters the Blood Temperature selection menu.
•
Cfg (
F4
): Enters the Configurations menu.
•
Esc
: Returns to the Main Screen.
CO Output Selection Menu
C O r a t e : 5 L p m
< 3 5 6 7 >
F1 F2
F3 F4
C O r a t e : 5 L p m
< 5 C a l >
F1
In the CO Rate menu, four CO rate wave types are available in
Liters per minute: 3, 5, 6, and 7. 5Cal is a 5Lpm square wave
calibration output. The current setting is displayed. Selecting a
new value returns you to the Main Screen. Press
Esc
to leave
the current setting unchanged.