Dry Bath N10 Series Operation manual
Chapter 4 Operation Guide
—
10
—
c)
Press “Stop” key without release, then press
key at the same time for 3 seconds, it will enter into
temperature calibration interface, see the right.
At this point,
“
” displayed, which indicates entering into external
temperature calibration program, the temperature displayed is external sensor
’s
current temperature and begins to rise to 40.0
℃
automatically.
It will automatically display
“
” six seconds
later, the upper displayed 32.0 is the current
temperature of the external sensor.
When the temperature reaches to 40
℃
constant
temperature, the decimal digit begins to flicker,
waiting for the calibrated value of 40
℃
. Read out
the actual value from the thermometer after 20
minutes.
Notes: Please read the actual value after 20
minutes’ constant temperature to ensure
the calibration accuracy.
Press
“Stop” key
to
view the
temperature
value of
external
sensor
,
if displayed 39.6
°
C
.
Press
“Stop” again, or wait for 6 seconds, then
displayed
“
”, modify the displayed
temperature to 39.6 by pressing
or
. Then press
to confirm the input value.
d) Then the instruments will heating to 100
℃
automatically
,
waiting till the instrument
is in 100
℃
thermostat, input temperature calibration value. After 20
minutes’
constant temperature, the user can read the actual value from thermometer.
Notes: Please read the actual value after 20
minutes’ constant temperature to ensure
the calibration accuracy.
If the actual readout of sensor is 101.5
℃
, modify
the temperature in display to 101.5 by pressing
or
. Then press “
” to confirm the input value.