Dry Bath N10 Series Operation manual
Chapter 4 Operation Guide
—
8
—
temperature in display is below 35
℃
. If the temperature is higher than 35
℃
, you
should wait until the temperature is below 35
℃
.
b) Inject olefin oil into one of the cone-shaped wells, and then put a thermometer into
this well (make sure that the precision of the thermometer should be within 0.1
℃
and the temperature ball should be absolutely immerged into the cone-shaped well).
See Fig a.
c)
Press “Stop” key without releasing, then press
key of temperature at the same time for 3 seconds,
it will enter into temperature calibration interface,
see the right. At this point,
“ ”displayed, which indicates entering into
temperature calibration program. The temperature displayed is current
temperature and begins to rise to 40.0
℃
automatically.
When the temperature reaches to 40
℃
constant
temperature, the decimal digit begins to flicker,
waiting for the calibrated value of 40
℃
. Read out the
actual value from the thermometer after 20 minutes.
Notes: Please read the actual value after 20
minutes’ constant temperature to ensure
the calibration accuracy.
Fig a
Thermometer
Olefin oil
Cone-shaped well