Dry Bath N10 Series Operation manual
Chapter 4 Operation Guide
—
9
—
If the actual read out of thermometer is 39.6
℃
,
Modify the temperature to 39.6 by pressing
or
. Then press
“Start” to confirm the input value.
d) Then the instrument will heat to 100
℃
automatically, waiting till the instrument is in
100
℃
thermostat, input temperature calibration value. After 20
minutes’ constant
temperature, the user can read the actual value from thermometer.
Notes: Please read the actual value after 20
minutes’ constant temperature to
ensure the calibration accuracy.
If the actual readout is 101.5
℃
, modify the
temperature in display to 101.5 by pressing
or
.
Then pres
s “Start” to confirm the input.
Notes: Pressing
“Stop” and “
” simultaneously during the temperature calibration
indicates exiting the temperature calibration program. The changed value will
be of no effect.
3.2. Calibrated with an external sensor, specific methods are as follows
:
Tips
:
External sensors are optional parts, it needs to be ordered separately.
a) After the startup of the instrument, it enters into waiting interface. Make sure that the
temperature in display is below 35
℃
. If the temperature is higher than 35
℃
, the user
should wait until the temperature is below 35
℃
.
b) Put one end of the external sensors into a block well, and insert the other end to the
interface. See Fig b.
Fig b