ERROR MESSAGES
1.
Hi - measured temperature too high, outside the measuring range
2.
Lo - too low a measured temperature, outside the measuring range
3.
Er1 - operating temperature outside the range 10~40
℃
4.
ErC - error occurs when data is read or written into the memory or the temperature correction is not completed
5.
When the battery voltage drops below 2.5V ± 0.1V, the low battery symbol appears in the display. Replace the
batteries.
6.
Blank screen:
a.
The thermometer switches off automatically after a short period of inactivity, this is a deliberate action of the
device - restart the device by pressing the power/measurement button
b.
Battery incorrectly installed - ensure battery terminals are facing the correct way
c.
Batteries flat - replace with new batteries
d.
Display does not start despite above instructions - contact service
06.
CLEANING AND STORAGE
1.
Do not touch or press the device sensor.
2.
To clean the thermometer remove the batteries and then clean the ear-tip with a soft cloth avoiding the temperature
sensor lens. The lens itself is cleaned with a cotton swab and the casing with a slightly damp, soft cloth.
3. Note:
Keep water away from the lens during the cleaning process. Otherwise, the lens may be damaged. The lens
may be scratched if cleaned with a hard object, which may cause inaccurate readings. Do not clean the thermometer
with corrosive cleaning agents. Do not immerse any part of the thermometer in liquid or allow liquid to penetrate the
thermometer during the cleaning process.
07.
TYPICAL HUMAN BODY TEMPERATURE
The human body is a complex biological system and the range of temperature that can be considered ‚normal’ depends in
large part on what part of the body we measure and on factors such as age, sex, skin colour and skin thickness. Women’s
body temperature is higher than men’s by about 0.3
℃
. In addition, body temperature in women increases by an additio
-
nal 0.3-0.5
℃
during ovulation.
08. SPECIFICATION
Place of measurement: forehead, ear, room, food, object | Units of measurement: degrees Celsius
(℃
) or degrees
Fahrenheit (°F) | Operating temperature: 10~40
℃
| Storage temperature: -20~50
℃
| Measurement distance: 1~3 cm |
Measurement range: forehead 22~43
℃
(71.6~109.4°F), ear 34~43
℃
(93.2~109.4°F), object 0~100
℃
(32~212°F) | Measu
-
rement accuracy: Forehead: ±0.2°C in the range 36.0
℃
-39.0
℃
,
±0.3°C in the ranges: 22.0
℃
-36.0
℃
and 39.0
℃
~43.0
℃
);
Ear: ±0.2°C in ranges 36.0
℃
-39.0
℃
,
±0.3°C in ranges: 34.0
℃
-36.0
℃
and 39.0
℃
~43.0
℃
; Subject/ear: ±1.0
℃
/±2.0
℉
|
Memory: 20 measurements | Dimensions: 145.5x45.4x43.4 mm | Weight (without batteries): 72g | Batteries: 2xAAA, DC
3V (not included) | Auto shut-off: after 10 seconds of inactivity
09. WARRANTY CARD
Dear customer, thank you for purchasing our Neno Medic T07 thermometer. If you have any problems operating the de
-
vice under normal conditions, please contact an authorized Neno brand service center or distributor. Keep your warranty
card in case of repair.
Product comes with a 24-month warranty. Warranty conditions can be found at:
https://neno.pl/gwarancja
Details, contact and service address can be found at:
https://neno.pl/kontakt
Specifications and kit contents are subject to change without notice. We apologise for any inconvenience.
KGK Trade declares that the Neno Medic T07 device complies with the essential requirements of Directive 2014/53/EU.
The full text of the declaration can be found at the link:
https://neno.pl/download/DOC/deklaracja-CE-Neno-Medic-T07.
10. MANUFACTURER’S DECLARATION EUT
Manufacturer’s guidelines and declaration - electromagnetic emissions - for all equipment and systems.
Guidelines and manufacturer's declaration - electromagnetic emissions
The infrared thermometer is intended for use in the electromagnetic environment specified below. The customer or user of the infrared thermometer
should ensure that it is used in such an environment.
Emissions test
Compatibility
Electromagnetic environment - guidance
RF emission CISPR 11
Group 1
The infrared thermometer uses RF energy only for its internal functions. Therefore, its RF emissions
are very low and are unlikely to cause any interference to nearby electronic equipment.
RF emission CISPR 11
Class B
The infrared thermometer is suitable for use in all establishments, including residential buildings and
those directly connected to the public low-voltage power supply network that supplies buildings used
for domestic purposes.
Guidelines and manufacturer's declaration - electromagnetic immunity
The infrared thermometer is intended for use in the electromagnetic environment specified below. The customer or user of the infrared thermometer
should ensure that it is used in such an environment.
Immunity test
Test level EN60601
Level of compliance
Electromagnetic environment - guidance
Electrostatic discharges
(ESD)
IEC 61000-4-2
±6kV contact
±8 kV air
±6kV contact
±8 kV air
Floors should be made of wood, concrete or ceramic tiles. If floors
are covered with synthetic material, the relative humidity should
be at least 30%.
Mains frequency
magnetic field (50/60 Hz)
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels characteristic
of a typical location in a typical commercial or hospital
environment.
Guidelines and manufacturer's declaration - electromagnetic immunity
The infrared thermometer is intended for use in the electromagnetic environment specified below. The customer or user of the infrared thermometer
should ensure that it is used in such an environment.
Immunity test
Test level EN60601
Level of compliance
Electromagnetic environment - guidance
RF radiation
IEC 61000-4-3
3 V/m
80 MHz
up to 2.5 GHz
3 V/m
Portable and mobile radio communication equipment should not be used
closer to any part of the thermometer, including cables, than the
recommended separation distance calculated from the equation applied
to the transmitter frequency. Recommended separation distance:
𝑑𝑑 = [
3,5
𝐸𝐸
1
]√𝑃𝑃
80MHz to 800 MHz
𝑑𝑑 = [
7
𝐸𝐸
1
]√𝑃𝑃
800 MHz to 2.5 GHz
where P is the maximum rated output power of the transmitter in Watts
(W) according to the transmitter manufacturer and d is the recommended
separation distance in metres (m). Field strengths from fixed RF
transmitters, as determined by an on-site electromagnetic test*, should
be less than the compliance level in each frequency range**. Interference
may occur in the vicinity of equipment marked with the following symbol:
NOTE:
At 80 MHz and 800 MHz a higher frequency band applies.
These guidelines may not apply in all situations. The electromagnetic field is affected by absorption and reflection from structures, objects and people
*. Field strengths from fixed transmitters such as base stations of radio telephones (mobile/wireless) and terrestrial mobile radios, amateur radio, AM
and FM radio transmissions and TV transmissions cannot theoretically be predicted with accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength at the location where the thermometer is
used exceeds the appropriate RF compliance level, observe the thermometer to verify proper operation. If abnormal operation is observed, additional
measures such as reorienting or moving the thermometer may be necessary.
**. In the frequency range 150 kHz to 80 MHz the field strength shall be less than 3 V/m.
Содержание Medic T07
Страница 1: ...Medic T07...
Страница 2: ...2 A B C A 1 A 2 A 3 A 6 A 4 A 7 A 5 A 8 D IP22...
Страница 57: ...PL DE CZ EN ET HU LT SK LV...
Страница 58: ...RO FI SE NO DK...
Страница 59: ......