Electromagnetic Compatibility Tables | 21
Electromagnetic Compatibility Tables
All medical electronic devices must conform to the requirements of IEC 60601-1-2. Precautions, adherences to
the Electromagnetic Compatibility (EMC) guideline information provided in this manual and verification of all
medical devices in simultaneous operation are required to ensure the electromagnetic compatibility and co-
existence of all other medical devices prior to a surgical procedure.
The following EMC tables are provided for reference:
Electromagnetic Emissions
,
Electromagnetic Immunity,
and
Recommended Separation Distances
.
Electromagnetic Emissions
Guidance and Manufacturer’s Declaration: Electromagnetic Emissions
This product is intended for use in the electromagnetic environment specified below. The customer or the user of the
product should assure that it is used in such an environment.
Emissions
Compliance
Electromagnetic Environment Guidance
RF emissions
CISPR 11
Group 1
The product uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The product is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic purposes.
Harmonic
emissions
IEC 61000-3-2
Not applicable
Voltage
fluctuations/
flicker emissions
IEC 61000-3-3
Not applicable