Guidance and manufacturer's declaration - electromagnetic immunity for all equipment and systems
ICS Chartr 200 VNG/ENG is intended for use in the electromagnetic environment specified below. The user of ICS Chartr
200 VNG/ENG should ensure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment -
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
+/- 6 kV contact
+/- 8 kV air
+/- 6 kV contact
+/- 8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Electrical fast tran-
sient/burst IEC 61000-4-
4
+/- 2 kV for power supply
lines
+/- 1 kV for input/output
lines
+/- 2 kV for power supply
lines
+/- 1 kV for input/output
lines
Mains power quality should be that of a
typical commercial or hospital envir-
onment.
Surge IEC 61000-4-5
+/- 1 kV line(s) to line(s)
+/- 2 kV line(s) to earth
+/- 1 kV line(s) to line(s)
+/- 2 kV line(s) to earth
Mains power quality should be that of a
typical commercial or hospital envir-
onment.
Voltage dips, short inter-
ruptions and voltage
variations on power sup-
ply input lines IEC
61000-4-11
<5 % U
T
(>95 % dip in U
T
)
for 0.5 cycle
40 % UT (60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5 s
<5 % U
T
(>95 % dip in U
T
)
for 0.5 cycle
40 % UT (60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5 s
Mains power quality should be that of a
typical commercial or hospital envir-
onment.
If the user of the ICS Chartr 200
VNG/ENG requires continued operation
during very long power mains inter-
ruptions, it is recommended that the ICS
Chartr 200 VNG/ENGbe powered from an
uninterruptible power supply or battery.
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
3 A/m
Swept Magnetic Fields
per AAMI
Power frequency magnetic fields should
be at levels characteristic of a typical loc-
ation in a typical commercial or hospital
environment.
Otometrics - ICS Chartr 200 VNG/ENG
65
11 Technical Specifications