9.2
Warning Notes
ICS Chartr 200 VNG/ENG Warning Notes
Equipment connected to the displayed connectors must be certified to relevant
EN/IEC safety standards, e.g., EN/IEC 60950. Mains connected equipment –
except EN/IEC 60601-1 certified equipment – must be powered from the Power-
tronix Isolation Station (X1ATWFHNOC1).
Equipment connected to the displayed connectors must be certified to relevant
EN/IEC safety standards, e.g., EN/IEC 60950. Mains connected equipment –
except EN/IEC 60601-1 certified equipment – must be powered from the Power-
tronix Isolation Station (X1ATWFHNOC1).
The ICS Chartr 200 should only be connected to power adapter type
FW7362M/15 from Friwo. For continued protection against fire hazard, replace
fuse with the same type and rating only.
Note 1:
There are no user-serviceable parts inside the ICS Chartr 200 cabinet. For the sake of safety, and in
order not to void the warranty, the cabinets should only be opened and serviced by authorized ser-
vice personnel. In case of defects, please make a detailed description of the defect(s) and contact
your supplier. Do not use a defective instrument.
Note 2:
Keep ICS Chartr 200 VNG/ENG away from liquids. Do not allow moisture inside the instrument.
Note 3:
Do not use the instrument in the presence of flammable anesthetics (gases).
Note 4:
Unwanted noise may occur if ICS Chartr 200 VNG/ENG is exposed to a strong radio field. Such noise
may interfere with the process of recording correct measurements. Many types of electrical devices,
e.g., mobile telephones, may generate radio fields. We recommend that the use of such devices in
the vicinity of ICS Chartr 200 VNG/ENG is restricted as much as possible.
Note 5:
It is recommended to install the unit in an environment that minimizes the amount of static elec-
tricity. For example, anti-static carpeting is recommended.
Note 6:
No parts may be eaten, burnt, or in any way used for purposes other than videonystagmography and
electronystagmography testing.
Note 7:
ICS Chartr VNG/ENG can be disposed of as normal electronic waste, according to local regulations.
Note 8:
For safety reasons, accessories connected to the equipment's outlet fittings must be identical to the
type supplied with the system.
Note 9:
To comply with EN 60601-1-1, the computer, printer, etc. must be connected to the isolation trans-
former.
Note 10:
Conductive parts with patient connection must not be in contact with other conductive parts at any
time. No defibrillators or HF surgical equipment should be applied to the patient when connected
to ICS Chartr 200 VNG/ENG at any time.
56
Otometrics - ICS Chartr 200 VNG/ENG
9 Safety