19 Technical specifications
132
MADSEN Zodiac
Operating environment
Temperature:
+15°C to +35°C (59°F to +95°F)
Air humidity:
10 to 90%, non-condensing
Air pressure:
60 kPa to 106 kPa
Warm-up time:
< 10 min. If stored in conditions not within specified operating environment
conditions, the device must warm up for 24 hours before being used.
Altitude correction
The admittance of a cavity depends on the atmospheric pressure. This means that when the atmospheric pressure changes,
the relation between mmho and ml changes. The following table can be used to calculate the difference.
Altitude (m)
Increase in mmho (%)
0
0
500
6
1000
13
1500
20
2000
27
2500
36
3000
45
Storage and handling
Temperature:
-20°C to +60°C (-4°F to +140°F)
Relative humidity:
10 % to 90 %, non-condensing
Air pressure:
50 kPa to 106 kPa
Dimensions (HxWxD)
Stand-alone version:
190 mm x 248 mm x 261 mm (7.5” x 9.8” x 10.3”)
PC-based version:
100 mm x 240 mm x 240 mm (3.9” x 9.4” x 9.4”)
Probe dimensions (HxWxD)
Quick Check probe:
28 mm x 22 mm x 125 mm (1.1” x 0.9” x 4.9”), Cable length 2800 mm (110”)
Diagnostic probe:
10 mm x 10 mm x 25 mm (0.4” x 0.4” x 1.0”), Cable length 2445 mm (96”)
Weight
Stand-alone version:
2.65 kg/5.85 lb
PC-based version:
1.65 kg/3.64 lb