Table 2 - Electromagnetic Immunity
Guidance and manufacturer’s declaration – electromagnetic immunity
The Dantec Clavis is intended for use in the electromagnetic environment specified below. The
customer or the user of the Dantec Clavis should assure that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment
- Guidance
Electrostatic Dis-
charge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
Complies
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30%.
Electrostatic fast
transient/burst
IEC 61000-4-4
±2 kV, 100Khz for
power supply lines
±1 kV, 100Khz for
input/output lines
Complies
Mains power quality should be that of a
typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV differential
mode
±2 kV common
mode
Complies
Mains power quality should be that of a
typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
<100% drop, 0/5
periods, 0°, 45°,
90°, 135°, 180°,
225°, 270°, 315°
100% dip, 1 period
30% dip, 25/30 peri-
ods
40% dip for 5 cycles
Complies
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the Dantec
Clavis requires continued operation
during power mains interruption, it is
recommended that the Dantec Clavis is
powered from a 9 Volt battery.
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
30 A/m
Complies
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.