Dantec Clavis User Guide
25
Table 4 - Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equip-
ment
Test
fre-
quency
(MHz)
Band
a)
(MHz)
Service
a)
Modulation
b)
Maximum
Power
(W)
Distance
(m)
IMMUNITY
TEST LEVEL
(V/m)
385
380 –390
TETRA 400
Pulse
modulation
b)
18 Hz
1,8
0,3
27
450
430 – 470
GMRS 460,
FRS 460
FM
c)
± 5 kHz
deviation
1 kHz sine
2
0,3
28
710
704 – 787
LTE Band 13,
17
Pulse
modulation
b)
217 Hz
0,2
0,3
9
745
780
810
800 – 960
GSM
800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation
b)
18 Hz
2
0,3
28
870
930
1,720
1,700 –
1,990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT; LTE
Band 1, 3, 4,
25; UMTS
Pulse
modulation
b)
217 Hz
2
0,3
28
1,845
1,970
2,450
2,400 –
2,570
Bluetooth,
WLAN,
802.11
b/g/n, RFID
2450, LTE
Band 7
Pulse
modulation
b)
217 Hz
2
0,3
28
5,240
5,100 –
5,800
WLAN
802.11 a/n
Pulse
modulation
b)
217 Hz
0,2
0,3
9
5,500
5,785
NOTE: If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna
and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC
61000-4-3.
a)
For some services, only the uplink frequencies are included.
b)
The carrier shall be modulated using a 50 % duty cycle square wave signal.
c)
As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does
not represent actual modulation, it would be worst case.
Declaration of Compliance for FCC
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to
part 15 of the FCC Rules.
Warning: Changes or modifications not expressly approved by the manufacturer could void the user's authority to
operate the equipment.