Dantec Clavis User Guide
15
Maintenance
Dantec Clavis requires no user maintenance other than cleaning the device after each use and replacing the
battery periodically.
Cleaning
The cleaning procedure must be in accordance with your local hygiene authority’s guidelines.
The user/operator shall clean device after every use.
1
Before you start cleaning the device, make sure it is switched off and that the electrode leads are
disconnected.
2
Wipe the device with a damp cloth.
3
Remove any excess of water with a dry cloth.
4
In case additional cleaning is required, gently wipe the device with a cloth moistened with a
maximum of 80% alcohol solution.
5
Always wipe the device dry after using the alcohol solution.
WARNING: Do not immerse the device in any liquid, or drip water into the connectors, or any other
openings in the cover.
CAUTION: Do not use solvent silicon based or abrasive cleaning agents to clean the device.
Replacing the Battery
Replace the battery when the indicator emits a yellow light.
Note that the device will automatically go into
“Fail Safe”
mode when the battery power is so low that it
would affect the performance.
CAUTION: Do not use rechargeable batteries. Use only a standard 9V alkaline battery – see the sec-
tion “Technical Data” in this user guide for further information.
CAUTION: Battery Leakage: If the unit is going to be stored for a prolonged period of time,
we recommend that the battery is removed to protect the unit from damage caused by battery
chemical leakage.
NOTE: Always follow the instructions accompanying the batteries.