1
ACNS Guidelines from the journal of clinical neurophysiology. 2011 December; Vol 28 No. 6.
2
Utility of prolonged bedside amplitude-integrated encephalogram in encephalopathic infants. Mathur AM, Morris LD, Teteh F, Inder TE, Zempel J. Am J
Perinatol. 2008 November; 25(10):611-5. Epub 2008 Oct 7.
3
Effects of treatment of subclinical neonatal seizures with aEEG: Randomize, Control Trial. Van Rooij LGM, Tost MC, Van Huffelen AC, et al. Pediatrics 2010;
125:e358-e366.
4
Atlas of amplitude integrated EEGs in the Newborn, 2nd Edition. Lena Hellström-Westas, Ingmar Rosen, Linda S. de Vries, pp. 81 and pp. 82.
5
Sleep-Wake Cycling on Amplitude-Integrated Electroencephalography in Term Newborns With Hypoxic-Ischemic Encephalopathy. Damjan Osredkar, MD,
Mona C. Toet, MD, Linda G. M. van Rooij, MD, Alexander C. van Huffelen, MD, PhD, Floris Groenendaal, MD, PhD, Linda S. de Vries, MD, PhD. Pediatrics
2005 February; Vol. 115 No. 2, pp. 327-332.
6
NeoReviews. Hellstrom-Westas, Rosen, deVries, Greisen. 2006 February; Vol 7 No. 2.
© 2012 Natus Medical Incorporated. All Rights Reserved. All product names appearing on this document are trademarks or registered trademarks owned,
licensed to, promoted or distributed by Natus Medical Incorporated, its subsidiaries or affiliates.
P/N 011272D
Natus Medical Incorporated
1501 Industrial Road
San Carlos, CA 94070 USA
1-800-303-0306
+1-650-802-0400
www.natus.com
Ordering information
Description
Catalog#
Olympic Brainz Monitor Kit - NA
(Includes: Monitor, roll stand, starter kit,
power cord, DAB and hard copy manuals)
OBM70001
Olympic Brainz Monitor Kit - EU
(Includes: Monitor, roll stand, starter kit,
power cord, DAB)
OBM70002
Olympic Brainz Monitor Kit - UK
(Includes: Monitor, roll stand, starter kit,
power cord, DAB)
OBM70003
Olympic Brainz Monitor Kit - NZ/AUS
(Includes: Monitor, roll stand, starter kit,
power cord, DAB)
OBM70004
Software options
RecogniZe Seizure Detection Software License Kit
OBM00092
(Not available
in the US)
Consumables
Neonatal Sensors – 12 sets
(1 set = 5 sensors) in a re-sealable pouch
OBM00042
Low Impedance needle electrodes - 6 sets
(1 set = 4 needles)
OBM00046
Wrap Hats (pack of 10 w/ dots)
OBM00043
Skin Markers (box of 10)
OBM00044
NuPrep Skin Preparation Gel - 4oz Tubes (3-pk)
102566N
Positioning Strips - Term and Pre-Term, pack of 20
(10 of each)
OBM00047
CFM Olympic Brainz Monitor
General specifications
Touch Screen Monitor
Weight
Dimensions
14.33 lbs (10 kg)
16.46 x 13.46 x 4.53 in
(418 x 342 x 115 mm)
Data Acquisition Box (DAB)
Weight
Dimensions
10 oz (280 g)
2.98 x 5.75 x 1.23 in
(75.7 x 146.1 x 31.2 mm)
Roll Stand
Weight
Dimensions
40 lbs (20 kg)
61.5 in height, 25 in base dia.
(1562 mm height, 635 mm base dia.)
Operation (all components)
Temperature
Relative humidity
Display
0 to 40 °C (32 to 104 °F)
25 to 90% at 40 °C (non-condensing)
Real-time EEG Waveform
Rapid pens (aEEG, Impedance - Computed
Rapid numeric (Impedance) - Computed
Histogram distribution over 15-second
intervals (aEEG, Impedance)
Color TFT LCD with resistive touchscreen,
15” (381 mm) diagonal, TFT color,
1024 x 768 pixel native resolution
Power supply (integrated)
Power supply unit
Power supply input voltage
EEG specifications sensitivity
Dynamic range
Update rate
Integrated AC, medical grade
100 - 240 VAC, 50/60 Hz, 4A - 2A
50 μVpk full scale maximum sensitivity
(< 1μV/mm)
0.30 - 10000 μVpp (1-20 Hz)
200 Hz (EEG Waveform)
Data Acquisition Box (DAB) Specifications
Differential channels
Frequency response
Analogue to digital converter
Sampling rate
Resolution
Sampling quantization
Input impedance (DC)
3
0.5 Hz ~ 450 Hz
SAR ADC (16x oversampling)
2000 Hz
16 bits
300 nV
>50 MΩ