Using the BRM2 Brain Monitor
BRM2 Brain Monitor Reference Manual |
29
Deleting files
To delete files from either internal storage or removable media:
1. Select the file you want to delete in either the
Internal Storage
or
Removable Media
file list.
2. Touch the appropriate
Delete
button.
Refreshing the file lists
To refresh both the
Internal Storage
and
Removable Media
file lists, touch
Refresh.
Restoration after AC supply interruption
If there is an AC supply interruption of one hour or less, the BRM2 Brain Monitor can resume the current
assessment and record data into the same data files when AC power is restored. The resulting data file
will have a blank period corresponding to the AC power interruption.
To continue the same assessment after an AC supply interruption:
1. When the AC supply is interrupted, turn off the BRM2 Brain Monitor’s AC Supply switch (see
Figure 8
on page 21).
2. After the AC supply has been restored, turn on the BRM2 Brain Monitor’s AC Supply switch and
log in as normal.
3. On
the
Main Menu
screen, touch
Assess Patient
.
4. When the message “Unexpected shutdown detected” displays, touch
Yes
to restore data from the
previous session. The previous assessment will resume.
If you touch
No
on the “Unexpected shutdown detected” dialog box, the previous assessment will be
closed and saved.