NI PXIe-4304/4305 and TB-4304 User Guide and Terminal Block Specifications
|
© National Instruments
|
5
Figure 2.
Installing NI PXIe-4304/4305
Modules
Connect the Signals
Caution
For EMC compliance, operate this product with shielded cables and
accessories.
To connect signals to the terminal block, refer to Figures 3 and 4 while completing the following
steps:
Note
You can find the pinout names and locations in Measurement & Automation
Explorer (MAX) at any time by right-clicking the device name under
Devices and
Interfaces
and selecting
Device Pinouts
.
1.
Loosen the captive top cover screws and remove the top cover.
2.
Loosen the strain-relief screws and remove the strain-relief bar.
3.
Prepare the shielded signal wire by stripping the insulation no more than 7 mm (0.28 in.).
4.
Run the shielded signal wires through the strain-relief opening.
1
Captive Screws
NI PXI