SCXI-1303 Terminal Block Calibration Procedure

2

ni.com

Equipment and Other Test Requirements

This section describes the equipment, software, documentation, and test 
conditions required for verifying the performance of your SCXI-1303 
terminal block.

Test Equipment

Verification requires a high-precision voltage source with at least 50 ppm 
accuracy, a multiranging 5 1/2 digit digital multimeter (DMM) with 
15 ppm accuracy, and a thermometer that is accurate to within 0.1 

°

C.

National Instruments recommends you use the following instruments for 
verifying the performance of your SCXI-1303 terminal block:

Calibrator—Fluke 5700A

DMM—NI 4060 or HP 34401A

If these instruments are not available, use the accuracy requirements listed 
above to select a substitute calibration standard.

Software and Documentation

You can find all the necessary information to verify the performance 
of the SCXI-1303 in this verification procedure. No other software or 
documentation is required. If you would like more information on the 
SCXI-1303, refer to the SCXI-1303 Terminal Block Installation Guide
which you can download from the National Instruments Web site at 

ni.com/manuals

Test Conditions

Follow these guidelines to optimize the connections and the environment 
during verification:

Keep connections to the SCXI-1303 terminal block short. Long cables 
and wires act as antennae, picking up extra noise that can affect 
measurements.

Use shielded copper wire for all cable connections to the device. 
Use twisted-pair wire to eliminate noise and thermal offsets.

Keep relative humidity below 80%.

Maintain a temperature between 15 and 35 

°

C.

Содержание SCXI-1303

Страница 1: ...terminal block Because these components are not user adjustable calibration consists of verification only without correcting for any error Verification is measuring the performance of a device and com...

Страница 2: ...quirements listed above to select a substitute calibration standard Software and Documentation You can find all the necessary information to verify the performance of the SCXI 1303 in this verificatio...

Страница 3: ...ignments Individual pins are identified by their column and row For example A3 denotes the terminal located in Column A and Row 3 This conforms to the labeling of the pins on the front connector of a...

Страница 4: ...tive input terminal of the calibrated DMM to the negative terminal of the 5 VDC power supply 3 Place the terminal block in a temperature controlled environment where the temperature is between 15 and...

Страница 5: ...e parts or modify equipment Return the terminal block to National Instruments to ensure that the safety features are not compromised If Tact Tmeas 0 5 C the terminal block temperature sensor is nonfun...

Страница 6: ...SCXI 1303 Terminal Block Calibration Procedure 6 ni com Figure 1 SCXI 1303 Front Connector Pin Assignments B1 C1 A1 5 VDC C32 B32 A32 A3 Temp Sensor A2 GND...

Страница 7: ...National Instruments Corporation 7 SCXI 1303 Terminal Block Calibration Procedure Figure 2 Location of Jumper W1 on the SCXI 1303 Terminal Block Jumper W1...

Отзывы: