SCXI-1303 Terminal Block Calibration Procedure

7.

Compare T

act

 to T

meas

.

If (T

meas

 

 0.5 

°

C) 

 T

act

 

 (T

meas

 

+

 0.5 

°

C), the performance of 

the terminal block temperature sensor has been verified.

If T

act 

<

 (T

meas

 

 0.5 

°

C),

 

the terminal block temperature sensor is 

nonfunctional. Do not substitute parts or modify equipment. 
Return the terminal block to National Instruments to ensure that 
the safety features are not compromised.

If T

act

 

>

 (T

meas

 

+

 0.5 

°

C), the terminal block temperature sensor is 

nonfunctional. Do not substitute parts or modify equipment. 
Return the terminal block to National Instruments to ensure that 
the safety features are not compromised.

8.

Return jumper W1 to its original position.

You have completed verifying the performance of the temperature sensor 
of your SCXI-1303 terminal block.

Содержание SCXI-1303

Страница 1: ...terminal block Because these components are not user adjustable calibration consists of verification only without correcting for any error Verification is measuring the performance of a device and com...

Страница 2: ...quirements listed above to select a substitute calibration standard Software and Documentation You can find all the necessary information to verify the performance of the SCXI 1303 in this verificatio...

Страница 3: ...ignments Individual pins are identified by their column and row For example A3 denotes the terminal located in Column A and Row 3 This conforms to the labeling of the pins on the front connector of a...

Страница 4: ...tive input terminal of the calibrated DMM to the negative terminal of the 5 VDC power supply 3 Place the terminal block in a temperature controlled environment where the temperature is between 15 and...

Страница 5: ...e parts or modify equipment Return the terminal block to National Instruments to ensure that the safety features are not compromised If Tact Tmeas 0 5 C the terminal block temperature sensor is nonfun...

Страница 6: ...SCXI 1303 Terminal Block Calibration Procedure 6 ni com Figure 1 SCXI 1303 Front Connector Pin Assignments B1 C1 A1 5 VDC C32 B32 A32 A3 Temp Sensor A2 GND...

Страница 7: ...National Instruments Corporation 7 SCXI 1303 Terminal Block Calibration Procedure Figure 2 Location of Jumper W1 on the SCXI 1303 Terminal Block Jumper W1...

Отзывы: