SCXI-1303 Terminal Block Calibration Procedure

Verification Procedure

This section contains step-by-step instructions for verifying the 
performance of the temperature sensor on your SCXI-1303 terminal block.

Verifying Temperature Sensor Performance

Complete the following steps to verify the performance of the temperature 
sensor on your terminal block:

1.

Connect a +5 VDC power source to the terminal block.

a.

Hold the terminal block vertically upright and view it from the 
rear. The terminals on the 96-pin DIN connector are designated 
as follows:

Column A is on the right, Column B is in the middle, and 
Column C is on the left.

Row 1 is at the bottom and Row 32 is at the top.

Figure 1 illustrates the connector pin assignments. Individual pins 
are identified by their column and row. For example, A3 denotes 
the terminal located in Column A and Row 3. This conforms to the 
labeling of the pins on the front connector of a mating SCXI 
module. It does not necessarily correspond to the labeling of the 
pins on the rear of the terminal block connector itself, which you 
can only view by opening the terminal block enclosure.

b.

Strip 0.5 inches of insulation from one end of a 22 AWG solid 
wire. Insert the stripped end of the wire into terminal A1 on the 
96-pin female DIN connector on the rear of the terminal block. 
Attach the other end of this wire to the positive terminal of the 
+5 VDC power supply.

c.

Strip 0.5 inches of insulation from one end of a 22 AWG solid 
wire. Insert the stripped end of the wire into terminal A2 on the 
96-pin female DIN connector on the rear of the terminal block. 
Attach the other end of this wire to the negative terminal of the 
+5 VDC power supply.

Содержание SCXI-1303

Страница 1: ...terminal block Because these components are not user adjustable calibration consists of verification only without correcting for any error Verification is measuring the performance of a device and com...

Страница 2: ...quirements listed above to select a substitute calibration standard Software and Documentation You can find all the necessary information to verify the performance of the SCXI 1303 in this verificatio...

Страница 3: ...ignments Individual pins are identified by their column and row For example A3 denotes the terminal located in Column A and Row 3 This conforms to the labeling of the pins on the front connector of a...

Страница 4: ...tive input terminal of the calibrated DMM to the negative terminal of the 5 VDC power supply 3 Place the terminal block in a temperature controlled environment where the temperature is between 15 and...

Страница 5: ...e parts or modify equipment Return the terminal block to National Instruments to ensure that the safety features are not compromised If Tact Tmeas 0 5 C the terminal block temperature sensor is nonfun...

Страница 6: ...SCXI 1303 Terminal Block Calibration Procedure 6 ni com Figure 1 SCXI 1303 Front Connector Pin Assignments B1 C1 A1 5 VDC C32 B32 A32 A3 Temp Sensor A2 GND...

Страница 7: ...National Instruments Corporation 7 SCXI 1303 Terminal Block Calibration Procedure Figure 2 Location of Jumper W1 on the SCXI 1303 Terminal Block Jumper W1...

Отзывы: