SCXI-1303 Terminal Block Calibration Procedure

4

ni.com

2.

Connect a calibrated DMM to the temperature-sensor output of the 
terminal block.

a.

Refer to Figure 2 to locate jumper W1 on your terminal block and 
verify that MTEMP is jumpered.

b.

Strip 0.5 inches of insulation from one end of a 22 AWG solid 
wire. Insert the stripped end of the wire into terminal A3 on the 
96-pin female DIN connector on the rear of the terminal block. 
Attach the other end of this wire to the positive input terminal of 
the calibrated DMM.

c.

Connect the negative input terminal of the calibrated DMM to the 
negative terminal of the +5 VDC power supply.

3.

Place the terminal block in a temperature-controlled environment 
where the temperature is between 15 and 35

°

C.

4.

When the terminal block temperature equilibrates with its 
surroundings, measure the temperature sensor output V

meas

 using 

a calibrated DMM.

5.

Measure the actual temperature T

act

 in the temperature-controlled 

environment using a calibrated thermometer.

6.

Convert V

meas

 (in volts) to measured temperature T

meas

 (in degrees

 

Celsius) by performing the following calculations:

a.

Calculate

b.

Calculate

c.

Calculate

where T

meas

 is in 

°

C

a = 1.295361 × 10

–3

b = 2.343159 × 10

–4

c = 1.018703 × 10

–7

x

2.5

V

me as

5000

---------------------------

=

y

V

m eas

x

-------------

ln

=

T

meas

1

a

y b

cy

2

+

(

)

+

----------------------------------

273.15

=

Содержание SCXI-1303

Страница 1: ...terminal block Because these components are not user adjustable calibration consists of verification only without correcting for any error Verification is measuring the performance of a device and com...

Страница 2: ...quirements listed above to select a substitute calibration standard Software and Documentation You can find all the necessary information to verify the performance of the SCXI 1303 in this verificatio...

Страница 3: ...ignments Individual pins are identified by their column and row For example A3 denotes the terminal located in Column A and Row 3 This conforms to the labeling of the pins on the front connector of a...

Страница 4: ...tive input terminal of the calibrated DMM to the negative terminal of the 5 VDC power supply 3 Place the terminal block in a temperature controlled environment where the temperature is between 15 and...

Страница 5: ...e parts or modify equipment Return the terminal block to National Instruments to ensure that the safety features are not compromised If Tact Tmeas 0 5 C the terminal block temperature sensor is nonfun...

Страница 6: ...SCXI 1303 Terminal Block Calibration Procedure 6 ni com Figure 1 SCXI 1303 Front Connector Pin Assignments B1 C1 A1 5 VDC C32 B32 A32 A3 Temp Sensor A2 GND...

Страница 7: ...National Instruments Corporation 7 SCXI 1303 Terminal Block Calibration Procedure Figure 2 Location of Jumper W1 on the SCXI 1303 Terminal Block Jumper W1...

Отзывы: