background image

© 2017–2019 National Instruments. All rights reserved. 

377026B-01

Aug19

Information is subject to change without notice. Refer to the 

NI Trademarks and Logo Guidelines 

at 

ni.com/trademarks

 for more 

information on NI trademarks. Other product and company names mentioned herein are trademarks or trade names of their respective 
companies. For patents covering NI products/technology, refer to the appropriate location: 

Help»Patents

 in your software, the 

patents.txt

 

file on your media, or the 

National Instruments Patents Notice

 at 

ni.com/patents

. You can find information about end-user license 

agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer to the 

Export Compliance Information 

at 

ni.com/legal/export-compliance

 for the NI global trade compliance policy and how to obtain relevant HTS codes, ECCNs, and 

other import/export data. NI MAKES NO EXPRESS OR IMPLIED WARRANTIES AS TO THE ACCURACY OF THE INFORMATION 
CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS. U.S. Government Customers: The data contained in this manual 
was developed at private expense and is subject to the applicable limited rights and restricted data rights as set forth in FAR 52.227-14, DFAR 
252.227-7014, and DFAR 252.227-7015.

Содержание SC Express PXIe-4309

Страница 1: ...tion 17 Specifications 17 Worldwide Support and Services 17 Software Calibrating the PXIe 4309 requires the installation of NI DAQmx on the calibration system Driver support for calibrating the PXIe 4...

Страница 2: ...d reference information PXIe 4309 Specifications PXIe 4309 specifications and calibration interval NI DAQmx Readme Operating system and application software support in NI DAQmx NI DAQmx Help Informati...

Страница 3: ...ut 0 1 V Output Impedance 50 DMM PXI 4071 or PXIe 4081 DC Voltage Input Range 10 V Accuracy 12 ppm of reading 0 5 ppm of range PXI Express Chassis PXIe 1062Q If this chassis is unavailable use another...

Страница 4: ...re information refer to the Maintain Forced Air Cooling Note to Users document available at ni com manuals Allow adequate warm up time for all of the instruments and equipment according to the manufac...

Страница 5: ...le for the calibration references Note The verification procedure requires performing a self calibration as a first step unless self verification is done immediately after an adjustment Refer to the S...

Страница 6: ...Wait ten minutes for the thermal EMF to stabilize iv Connect the double banana plug shorting bar to the input of the CAL 4309 as shown in Figure 1 v Wait two minutes for the thermal EMF to stabilize F...

Страница 7: ...p and clear the task g Compare the averaged values to the specified test limits in Table 2 1 TB 4309 ST Terminal Block 2 Aln Input Short C COPYRIGHT 2016 R FOR PATENTS NI COM PATENTS TB 4309 A1 B1 C1...

Страница 8: ...bled 5 5 1 0 V 10 V 50 50 15 V 75 75 AI0 31 0 1 V Every Sample Disabled 4 4 1 0 V 10 V 15 V AI0 15 0 1 V None Enabled 2 2 1 0 V 10 V 15 V Verification of channels 0 through 31 provides coverages of ea...

Страница 9: ...the PXIe 4309 to the calibrator output i Attach the CAL 4309 terminal block to the PXIe 4309 module ii Tighten the mounting screw at the top of the terminal block to secure the terminal block to the m...

Страница 10: ...air nickel plated copper wires to AI0 through AI7 to allow the connection of the PXIe 4309 inputs to the calibrator one channel at a time iii Reinstall the top cover and tighten the top cover screws i...

Страница 11: ...oltage channel task using the values in Table 5 c Start the task d Drop the first sample acquired for each channel e Average the remaining samples on a per channel basis f Stop and clear the task g Co...

Страница 12: ...0 9999357 1 0000643 0 9999627 1 0000373 1 0 V 1 0000643 0 9999357 1 0000373 0 9999627 10 V 10 V 9 999445 10 000555 9 999715 10 000285 10 V 10 000555 9 999445 10 000285 9 999715 15 V 15 V 14 999019 15...

Страница 13: ...ns required to adjust the PXIe 4309 Depending on the terminal block used for verification refer to step a or b for specifics about how to connect the calibrator output to the PXIe 4309 input 1 Measure...

Страница 14: ...M input as shown in Figure 6 i Loosen the top cover screws and remove the top cover ii Attach the TB 4309 ST terminal block to the PXIe 4309 module iii Tighten the mounting screw at the top of the ter...

Страница 15: ...pedance 10 G Auto Zero On ADC Calibration On DC Noise rejection High Order C COPYRIGHT 2016 R FOR PATENTS NI COM PATENTS TB 4309 A1 B1 C1 J12 AI28 AI28 AI12 AI12 AI20 AI20 AI4 AI4 AI29 AI29 AI13 AI13...

Страница 16: ...mxAdjust4309Cal and enter the measured internal reference value 4 Call DAQmx Close External Calibration DAQmxCloseExtCal to close the session 5 Use the action cancel if there has been any error during...

Страница 17: ...NI Factory Installation Services repairs extended warranty and other services Visit ni com register to register your NI product Product registration facilitates technical support and ensures that you...

Страница 18: ...at ni com patents You can find information about end user license agreements EULAs and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at n...

Отзывы: