PXIe-4309 Calibration Procedure
|
© National Instruments
|
7
Figure 2.
PXIe-4309 DC Offset Verification Setup Using the TB-4309 (ST)
2.
Verify the accuracy for each specified configuration in Table 2.
a.
Configure an AI voltage channel task using the values in Table 3.
b.
Start the task.
c.
Acquire the samples with the PXIe-4309.
d.
Drop the first sample acquired for each channel.
e.
Average the remaining samples on a per channel basis.
f.
Stop and clear the task.
g.
Compare the averaged values to the specified test limits in Table 2.