PXIe-4309 Calibration Procedure

|

© National Instruments

|

3

Test Equipment

Table lists the equipment recommended for the performance verification and adjustment 
procedures of the PXIe-4309. If the recommended equipment is not available, select a substitute 
using the requirements listed in Table 1.

Table 1.  

Recommended Equipment for PXIe-4309 Verification and Adjustment

Equipment

Recommended Model

Requirements

Calibrator

Fluke 5700A

Output Range:
±15 V, ±10 V, ±1 V and ±0.1 V

Accuracy:
10 ppm of output
(±15 V, ±10 V, ±1 V)

20 ppm of output
(±0.1 V)

Output Impedance: 

50 

.

DMM

PXI-4071 or PXIe-4081

DC Voltage

Input Range: 10 V

Accuracy:
12 ppm of reading,
0.5 ppm of range

PXI Express Chassis

PXIe-1062Q

If this chassis is unavailable, use 
another PXI Express chassis, such as 
the PXIe-1082 or PXIe-1078.

Connection Accessory

CAL-4309

For automated verification use the 
CAL-4309.

For manual verification use the 
TB-4309 (ST) or TB-4309 (MT). 
When using these terminal blocks for 
verification connect only one 
channel to the calibrator at a time.

Low thermal EMF test lead kit

Pomona Model 72928

Low thermal EMF banana-to-banana 
and banana-to-probe test lead sets.

Low thermal EMF double 
banana plug shorting bar

Pomona Model 5145

Low thermal EMF double banana 
plug shorting bar.

Содержание SC Express PXIe-4309

Страница 1: ...tion 17 Specifications 17 Worldwide Support and Services 17 Software Calibrating the PXIe 4309 requires the installation of NI DAQmx on the calibration system Driver support for calibrating the PXIe 4...

Страница 2: ...d reference information PXIe 4309 Specifications PXIe 4309 specifications and calibration interval NI DAQmx Readme Operating system and application software support in NI DAQmx NI DAQmx Help Informati...

Страница 3: ...ut 0 1 V Output Impedance 50 DMM PXI 4071 or PXIe 4081 DC Voltage Input Range 10 V Accuracy 12 ppm of reading 0 5 ppm of range PXI Express Chassis PXIe 1062Q If this chassis is unavailable use another...

Страница 4: ...re information refer to the Maintain Forced Air Cooling Note to Users document available at ni com manuals Allow adequate warm up time for all of the instruments and equipment according to the manufac...

Страница 5: ...le for the calibration references Note The verification procedure requires performing a self calibration as a first step unless self verification is done immediately after an adjustment Refer to the S...

Страница 6: ...Wait ten minutes for the thermal EMF to stabilize iv Connect the double banana plug shorting bar to the input of the CAL 4309 as shown in Figure 1 v Wait two minutes for the thermal EMF to stabilize F...

Страница 7: ...p and clear the task g Compare the averaged values to the specified test limits in Table 2 1 TB 4309 ST Terminal Block 2 Aln Input Short C COPYRIGHT 2016 R FOR PATENTS NI COM PATENTS TB 4309 A1 B1 C1...

Страница 8: ...bled 5 5 1 0 V 10 V 50 50 15 V 75 75 AI0 31 0 1 V Every Sample Disabled 4 4 1 0 V 10 V 15 V AI0 15 0 1 V None Enabled 2 2 1 0 V 10 V 15 V Verification of channels 0 through 31 provides coverages of ea...

Страница 9: ...the PXIe 4309 to the calibrator output i Attach the CAL 4309 terminal block to the PXIe 4309 module ii Tighten the mounting screw at the top of the terminal block to secure the terminal block to the m...

Страница 10: ...air nickel plated copper wires to AI0 through AI7 to allow the connection of the PXIe 4309 inputs to the calibrator one channel at a time iii Reinstall the top cover and tighten the top cover screws i...

Страница 11: ...oltage channel task using the values in Table 5 c Start the task d Drop the first sample acquired for each channel e Average the remaining samples on a per channel basis f Stop and clear the task g Co...

Страница 12: ...0 9999357 1 0000643 0 9999627 1 0000373 1 0 V 1 0000643 0 9999357 1 0000373 0 9999627 10 V 10 V 9 999445 10 000555 9 999715 10 000285 10 V 10 000555 9 999445 10 000285 9 999715 15 V 15 V 14 999019 15...

Страница 13: ...ns required to adjust the PXIe 4309 Depending on the terminal block used for verification refer to step a or b for specifics about how to connect the calibrator output to the PXIe 4309 input 1 Measure...

Страница 14: ...M input as shown in Figure 6 i Loosen the top cover screws and remove the top cover ii Attach the TB 4309 ST terminal block to the PXIe 4309 module iii Tighten the mounting screw at the top of the ter...

Страница 15: ...pedance 10 G Auto Zero On ADC Calibration On DC Noise rejection High Order C COPYRIGHT 2016 R FOR PATENTS NI COM PATENTS TB 4309 A1 B1 C1 J12 AI28 AI28 AI12 AI12 AI20 AI20 AI4 AI4 AI29 AI29 AI13 AI13...

Страница 16: ...mxAdjust4309Cal and enter the measured internal reference value 4 Call DAQmx Close External Calibration DAQmxCloseExtCal to close the session 5 Use the action cancel if there has been any error during...

Страница 17: ...NI Factory Installation Services repairs extended warranty and other services Visit ni com register to register your NI product Product registration facilitates technical support and ensures that you...

Страница 18: ...at ni com patents You can find information about end user license agreements EULAs and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at n...

Отзывы: