PXIe-4309 Calibration Procedure
|
© National Instruments
|
3
Test Equipment
Table 1 lists the equipment recommended for the performance verification and adjustment
procedures of the PXIe-4309. If the recommended equipment is not available, select a substitute
using the requirements listed in Table 1.
Table 1.
Recommended Equipment for PXIe-4309 Verification and Adjustment
Equipment
Recommended Model
Requirements
Calibrator
Fluke 5700A
Output Range:
±15 V, ±10 V, ±1 V and ±0.1 V
Accuracy:
10 ppm of output
(±15 V, ±10 V, ±1 V)
20 ppm of output
(±0.1 V)
Output Impedance:
50
Ω
.
DMM
PXI-4071 or PXIe-4081
DC Voltage
Input Range: 10 V
Accuracy:
12 ppm of reading,
0.5 ppm of range
PXI Express Chassis
PXIe-1062Q
If this chassis is unavailable, use
another PXI Express chassis, such as
the PXIe-1082 or PXIe-1078.
Connection Accessory
CAL-4309
For automated verification use the
CAL-4309.
For manual verification use the
TB-4309 (ST) or TB-4309 (MT).
When using these terminal blocks for
verification connect only one
channel to the calibrator at a time.
Low thermal EMF test lead kit
Pomona Model 72928
Low thermal EMF banana-to-banana
and banana-to-probe test lead sets.
Low thermal EMF double
banana plug shorting bar
Pomona Model 5145
Low thermal EMF double banana
plug shorting bar.