PXIe-4309 Calibration Procedure
Table 4.
DC Voltage Verification Test Limits
PXIe-4309
Configuration
Calibrator
Settings
As-Found Limits (V)
As-Left Limits (V)
Channels
*
Input
Range
DCV
Amplitude
Lower
Upper
Lower
Upper
AI0..7
†
±0.1 V
+0.1 V
0.0999893
0.1000107
0.0999920
0.1000080
-0.1 V
-0.1000107
-0.0999893
-0.1000080
-0.0999920
±1.0 V
+1.0 V
0.9999357
1.0000643
0.9999627
1.0000373
-1.0 V
-1.0000643
-0.9999357
-1.0000373
-0.9999627
±10 V
+10 V
9.999445
10.000555
9.999715
10.000285
-10 V
-10.000555
-9.999445
-10.000285
-9.999715
±15 V
+15 V
14.999019
15.000981
14.999424
15.000576
-15 V
-15.000981
-14.999019
-15.000576
-14.999424
*When using the TB-4309 (ST) or TB-4309 (MT) to perform the DC voltage verification, connect only
one channel to the calibrator at a time.
†Verification of channels 0 through 7 provides coverage of each of the 8 independent signal
conditioning paths and ADCs of the PXIe-4309.