4
|
ni
.c
o
m
|
NI
P
X
Ie
-56
94 Calibr
a
tion Pro
c
e
dure
IF digitizer
NI PXIe-5622
Adjusting IF gain for bypass path
Adjusting IF gain
Adjusting IF flatness
198941D-0XL
Signal generators
(quantity: 2,
RF Source 1
and
RF Source 2
)
Anritsu MG3692C
Options 2A, 3, 4, 15A,
and 22
Characterizing RF source power
Characterizing cable loss
Verifying third-order intercept point
out-of-band
Adjusting IF gain for bypass path
Adjusting IF gain
Adjusting IF flatness
Frequency range: 8 MHz to 270 MHz
Leveled power: -115 dBm to 18 dBm
Power accuracy: ±1.5 dB
Harmonics (typical):
0.1 MHz to
≤
10 MHz ...........<-30 dBc
>10 MHz to
≤
100 MHz ........<-40 dBc
>100 MHz to
≤
2.2 GHz........<-50 dBc
>2.2 GHz to
≤
20 GHz ..........<-30 dBc
Nonharmonic spurious:
0.1 MHz to
≤
10 MHz ...........<-30 dBc
>10 MHz to
≤
2.2 GHz..........<-60 dBc
>2.2 GHz to
≤
20 GHz ..........<-60 dBc
Output VSWR: <2.0:1 (typical)
Spectrum analyzer
Rohde & Schwarz
FSU26
Characterizing RF source power
Verifying third-order intercept point
out-of-band
Frequency range: 10 MHz to 3.6 GHz
Noise floor: <-152 dBm/Hz
Third-order intercept point (entire frequency range):
>10 dBm
SMA (m)-to-SMA (m)
semi flexible cable
NI 190412B-04
Characterizing RF source power
Characterizing cable loss
—
Table 1.
Recommended Equipment for NI 5694 Calibration (Continued)
Equipment
Recommended
Model
Where Used
Minimum Requirements