2
|
ni.com
|
NI PXIe-5694 Calibration Procedure
Software
Calibrating the NI 5694 requires you to install the following software on the calibration system:
•
NI-RFSA version 14.1 or later
•
NI Spectral Measurements Toolkit version 2.6 or later
You can download all required software from
ni.com/downloads
.
NI-RFSA supports programming the calibration procedures in the LabVIEW, C, and
LabWindows
™
/CVI
™
application development environments (ADEs). When you install
NI-RFSA, you need to install support only for the ADE that you intend to use.
Documentation
You might find the following documents helpful as you perform the calibration procedure:
•
NI PXIe-5694 Getting Started Guide
•
NI PXIe-5694 Specifications
•
NI RF Vector Signal Analyzers Help
The latest versions of these documents are available on
ni.com/manuals
.
Test Equipment
Table 1 lists the equipment NI recommends for the performance verification and adjustment
procedures. If the recommended equipment is not available, select a substitute using the
minimum requirements listed in the table.