NI PXIe-5694 Calibration Procedure
|
© National Instruments
|
19
5.
Measure the channel B power using the appropriate sensor calibration factor for the
configured frequency. Record this value in the corresponding
Reference Source
Measurement
cell.
6.
Repeat steps 4 and 5 for each frequency listed in Table 5.
7.
Disconnect power sensor B from the RF source 1 output.
8.
Connect the 3.5mm (m)-to-3.5mm (m) digitizer cable to the RF source 1 output.
254
255.6
257.2
258.8
260.4
262
263.6
265.2
266.8
268.4
Table 5.
Digitizer Cable Loss (Continued)
Frequency
(MHz)
Reference
Source
Measurement
Digitizer Cable
Loss
Splitter
Tracking Error
(dB)
RF Source
Cable / Splitter
Loss (dB)