30
|
ni.com
|
NI PXIe-5694 Calibration Procedure
16. Repeat steps 4 through 18 with the new settings and the following changes:
•
When configuring RF source 1 and 2 output power, use a device under test (DUT)
input power of -5 dBm instead of -0 dBm.
•
Record your results using Table 10.
17. Repeat steps 4 through 18 with the following changes:
•
When configuring RF source 1 and 2 output power, use a device under test (DUT)
input power of -12 dBm instead of -0 dBm.
•
Reconfigure the NI 5694 to enable step gain.
•
Record your results using Table 11.
Table 10.
TOI-OB Calculation,
DUT Input Power -5 dBm (per Tone), Downconversion Enabled
Device
Instantaneous
Bandwidth (Path)
RF Source 1
Frequency
(MHz)
RF Source 2
Frequency
(MHz)
IF OUT
Power
(dBm)
TOI
Out-of-Band
(dBm)
5 MHz
201.1
208.6
186.1
178.6
1.4 MHz
195.7
197.8
191.5
189.4
400 kHz
194.2
194.8
193
192.4
110 kHz
193.765
193.93
193.435
193.27
30 kHz
193.645
193.69
193.555
193.51
Table 11.
TOI-OB Calculation,
DUT Input Power -12 dBm (per Tone), Downconversion Enabled
Device
Instantaneous
Bandwidth (Path)
RF Source 1
Frequency
(MHz)
RF Source 2
Frequency
(MHz)
IF OUT
Power
(dBm)
TOI-OB
(dBm)
5 MHz
201.1
208.6
186.1
178.6
1.4 MHz
195.7
197.8
191.5
189.4