Figure 11. Frequency Reference Accuracy Adjustment Equipment Setup (Reference
Connections Not Shown)
250 kHz–20 GHz Signal Generator
ACCESS
ACTIVE
REF IN
1–20 MHz
±10 dBm NOM
17 dBm MAX
PULSE
IN
–0.5V MIN
+5.5V MAX
FM IN
±1 VDC NOM
±2 VDC MAX
AM IN
±1 VDC NOM
±2 VDC MAX
ALC IN
±15 VDC MAX
RF OUT
250 kHz –
20 GHz
REF OUT
10 MHz
5 dBm ±2 dBm NOM
REF OUT 2
100 MHz
5 dBm ±2 dBm NOM
TRIG
IN/OUT
3.3V CMOS
IN/OUT
–0.5V MIN IN
5.5V MAX IN
ESD
SENSITIVE
ALL PORTS
50
Ω
1
2
3
1. PXIe-5654 RF Analog Signal Generator
2. Signal Source Analyzer
3. K(m)-to-K(m) Cable
2.
Configure the signal source analyzer using the following settings:
•
Center frequency: 10 GHz
•
Reference level: 10 dBm
•
Frequency counter resolution: 0.1 Hz
•
Reference Clock source: External
3.
Call the niRFSG Initialize External Calibration VI using the appropriate password.
Note
The default password for password-protected operations is
NI
.
4.
Call the niRFSG 5654 OCXO Cal Initialize VI.
5.
Call the niRFSG 5654 OCXO Cal Configure VI.
6.
Use the signal source analyzer to measure the frequency at the peak of the signal.
7.
Use the frequency measured in step 6 as the value of the
measured frequency
input of
the niRFSG 5654 OCXO Cal Adjust VI.
28
|
ni.com
|
PXIe-5654 Calibration Procedure