Figure 10. PULSE IN Operation Verification Equipment Setup
250 kHz–20 GHz Signal Generator
ACCESS
ACTIVE
REF IN
1–20 MHz
±10 dBm NOM
17 dBm MAX
PULSE
IN
–0.5V MIN
+5.5V MAX
FM IN
±1 VDC NOM
±2 VDC MAX
AM IN
±1 VDC NOM
±2 VDC MAX
ALC IN
±15 VDC MAX
RF OUT
250 kHz –
20 GHz
REF OUT
10 MHz
5 dBm ±2 dBm NOM
REF OUT 2
100 MHz
5 dBm ±2 dBm NOM
TRIG
IN/OUT
3.3V CMOS
IN/OUT
–0.5V MIN IN
5.5V MAX IN
ESD
SENSITIVE
ALL PORTS
50
Ω
2
1
NI USB-6501
24-line Digital I/O
5
3
4
1. USB-6501 Digital I/O Device
2. BNC(m)-to-SMB(f) Cable
3. PXIe-5654 RF Analog Signal Generator
4. K(m)-to-K(m) Cable
5. Signal Source Analyzer
3.
Create a new device session for the PXIe-5654.
4.
Configure the signal source analyzer using the following settings:
•
Reference level: 5 dBm
•
Start frequency: 0.25 MHz
•
Stop frequency: 10,000 MHz
•
Span (kHz): 2 KHz
•
Resolution bandwidth: 30 Hz
•
Video bandwidth: 100 Hz
•
RF attenuation: 30 dB
5.
Configure the PXIe-5654 using the following settings:
•
Pulse Modulation Enabled: TRUE
•
Frequency: 5 GHz
•
Power Level: 10 dBm
6.
Commit the settings to hardware.
7.
Initiate signal generation.
8.
Enable the RF output.
26
|
ni.com
|
PXIe-5654 Calibration Procedure