Table 1. Recommended Equipment for PXIe-5654 Calibration (Continued)
Equipment
Recommended
Model
Where Used
Minimum
Requirements
Power meter
Anritsu ML2438A
with a SC7413A
Power Sensor
Verifications:
•
10 MHz reference
amplitude accuracy
•
100 MHz reference
amplitude accuracy
•
RF OUT amplitude
accuracy
•
RF OUT maximum
power
•
RF OUT minimum
power
Adjustments:
•
RF OUT power
Range: -15 dBm to
20 dBm
Frequency range:
250 kHz to 20 GHz
Accuracy: <±4.0%
4
|
ni.com
|
PXIe-5654 Calibration Procedure