Verifying RF OUT Phase Noise
This procedure verifies that the internal frequency generation circuitry is adjusted for correct
phase noise accuracy.
1.
Connect the PXIe-5654 RF OUT front panel connector to the signal source analyzer RF
input connector using the K(m)-to-K(m) cable as shown in the following figure.
Figure 6. RF OUT Phase Noise Verification Equipment Setup (Reference Connections
Not Shown)
250 kHz–20 GHz Signal Generator
ACCESS
ACTIVE
REF IN
1–20 MHz
±10 dBm NOM
17 dBm MAX
PULSE
IN
–0.5V MIN
+5.5V MAX
FM IN
±1 VDC NOM
±2 VDC MAX
AM IN
±1 VDC NOM
±2 VDC MAX
ALC IN
±15 VDC MAX
RF OUT
250 kHz –
20 GHz
REF OUT
10 MHz
5 dBm ±2 dBm NOM
REF OUT 2
100 MHz
5 dBm ±2 dBm NOM
TRIG
IN/OUT
3.3V CMOS
IN/OUT
–0.5V MIN IN
5.5V MAX IN
ESD
SENSITIVE
ALL PORTS
50
Ω
1
2
3
1. PXIe-5654 RF Analog Signal Generator
2. Signal Source Analyzer
3. K(m)-to-K(m) Cable
2.
Create a new device session for the PXIe-5654.
3.
Configure the signal source analyzer using the following settings:
•
Phase noise measurement using cross-correlation phase-locked loops (PLLs) and
internal generators
•
Automatic device under test (DUT) detection
•
Spur suppression enabled
PXIe-5654 Calibration Procedure
|
© National Instruments
|