NI PXI-4070 Calibration Procedure
56
ni.com
21. Call
niDMM_CalAdjustGain
with the following parameters:
•
Mode
=
NIDMM_VAL_DC_CURRENT
•
Range
=
1
•
Input Resistance
=
NIDMM_VAL_RESISTANCE_NA
•
Expected Value
=
1
22. Output –1 A on the calibrator with the current output set to
AUX
.
23. Call
niDMM_CalAdjustGain
with the following parameters:
•
Mode
=
NIDMM_VAL_DC_CURRENT
•
Range
=
1
•
Input Resistance
=
NIDMM_VAL_RESISTANCE_NA
•
Expected Value
=
–1
24. Output 0 A on the calibrator with the current output set to
AUX
.
25. Call
niDMM_CalAdjustOffset
with the following parameters:
•
Mode
=
NIDMM_VAL_DC_CURRENT
•
Range
=
1
•
Input Resistance
=
NIDMM_VAL_RESISTANCE_NA
26. Call
niDMM_CalAdjustOffset
with the following parameters:
•
Mode
=
NIDMM_VAL_AC_CURRENT
•
Range
=
1
•
Input Resistance
=
NIDMM_VAL_RESISTANCE_NA
27. Call
niDMM_CalAdjustMisc
with
Type
set to
NIDMM_EXTCAL_MISCCAL_SECTION
.
You have completed adjusting the current modes of the NI 4070. Select one
of the following options:
•
To verify that the NI 4070 is now operating within its specifications,
go to the
section and complete the appropriate
procedures.
•
To finish the calibration and close the session, go to the
section.