National Instruments Corporation
41
NI PXI-4070 Calibration Procedure
3.
Call
niDMM_InitExtCal
with the resource descriptor of the NI 4070
and your valid user password to output a calibration session
(
Cal Session
) that you can use to perform NI-DMM calibration or
regular measurement functions.
Note
You will use
Cal Session
in all subsequent function calls.
Note
The default user password for adjusting the NI 4070 is
NI
.
4.
Call
niDMM_ConfigurePowerLineFrequency
with
PowerLine
Frequency
set to
50
or
60
, depending on the power line frequency
(in Hz) that your instruments are powered from; select
50
for 400 Hz
power line frequencies
5.
Output 100 mV on the calibrator with the range locked to 2.2 V.
6.
Call
niDMM_CalAdjustGain
with the following parameters:
•
Mode
=
NIDMM_VAL_DC_VOLTS
•
Range
=
0.1
•
Input Resistance
=
NIDMM_VAL_10_MEGAOHM
•
Expected Measurement
=
0.1
7.
Output –100 mV on the calibrator with the range locked to 2.2 V.
8.
Call
niDMM_CalAdjustGain
with the following parameters:
•
Mode
=
NIDMM_VAL_DC_VOLTS
•
Range
=
0.1
•
Input Resistance
=
NIDMM_VAL_10_MEGAOHM
•
Expected Measurement
=
–0.1
9.
Output 10 V on the calibrator.
10. Call
niDMM_CalAdjustGain
with the following parameters:
•
Mode
=
NIDMM_VAL_DC_VOLTS
•
Range
=
10
•
Input Resistance
=
NIDMM_VAL_GREATER_THAN_10_GIGAOHM
•
Expected Measurement
=
10
11. Output –10 V on the calibrator.
12. Call
niDMM_CalAdjustGain
with the following parameters:
•
Mode
=
NIDMM_VAL_DC_VOLTS
•
Range
=
10
•
Input Resistance
=
NIDMM_VAL_GREATER_THAN_10_GIGAOHM
•
Expected Measurement
=
–10