NI PXI-4070 Calibration Procedure
16
ni.com
38. Output –100 mV on the calibrator with the range locked to 2.2 V. This
range prevents a 50
Ω
calibrator output resistance from creating a
voltage divider with the internal resistance of the NI 4070.
39. Set the input resistance of the NI 4070 to >10 G
Ω
by calling
niDMM_SetAttributeViReal64
with the following parameters:
•
Attribute_ID
=
NIDMM_ATTR_INPUT_RESISTANCE
•
Attribute_Value
=
NIDMM_VAL_GREATER_THAN_10_GIGAOHM
40. Call
niDMM_Read
. Subtract the previously stored 100 mV >10 G
Ω
mode offset from this measurement and verify that the result falls
between the limits listed in Table 11.
41. Set the input resistance of the NI 4070 to 10 M
Ω
by calling
niDMM_SetAttributeViReal64
with the following parameters:
•
Attribute_ID
=
NIDMM_ATTR_INPUT_RESISTANCE
•
Attribute_Value
=
NIDMM_VAL_10_MEGAOHM
42. Call
niDMM_Read
. Subtract the previously stored 100 mV >10 M
Ω
mode offset from this measurement and verify that the result falls
between the limits listed in Table 11.
43. Output 1 V on the calibrator.
44. Call
niDMM_ConfigureMeasurement
with the following
parameters:
•
Function
=
NIDMM_VAL_DC_VOLTS
•
Range
=
1
•
Resolution
=
1e–6
45. Set the input resistance of the NI 4070 to >10 G
Ω
by calling
niDMM_SetAttributeViReal64
with the following parameters:
•
Attribute_ID
=
NIDMM_ATTR_INPUT_RESISTANCE
•
Attribute_Value
=
NIDMM_VAL_GREATER_THAN_10_GIGAOHM
46. Call
niDMM_Read
. Verify that this measurement falls between the
limits listed in Table 11.
47. Set the input resistance of the NI 4070 to 10 M
Ω
by calling
niDMM_SetAttributeViReal64
with the following parameters:
•
Attribute_ID
=
NIDMM_ATTR_INPUT_RESISTANCE
•
Attribute_Value
=
NIDMM_VAL_10_MEGAOHM
48. Call
niDMM_Read
. Verify that this measurement falls between the
limits listed in Table 11.
49. Output –1 V on the calibrator.