NI PXI-4070 Calibration Procedure
32
ni.com
25. Call
niDMM_ConfigureMeasurement
with the following
parameters:
•
Function
=
NIDMM_VAL_2_WIRE_RES
•
Range
=
100e6
•
Resolution
=
100
26. Call
niDMM_Read
. Subtract the previously stored 100 M
Ω
range offset
from this measurement. Verify that the result falls between the
tolerances listed in Table 14.
27. Output 10 M
Ω
on the calibrator with no external sense or 2-wire
compensation.
28. Call
niDMM_ConfigureMeasurement
with the following
parameters:
•
Function
=
NIDMM_VAL_2_WIRE_RES
•
Range
=
10e6
•
Resolution
=
10
29. Call
niDMM_Read
. Subtract the previously stored 10 M
Ω
range offset
from this measurement. Verify that the result falls between the
tolerances listed in Table 14.
30. Output 1 M
Ω
on the calibrator with no external sense or 2-wire
compensation.
31. Call
niDMM_ConfigureMeasurement
with the following
parameters:
•
Function
=
NIDMM_VAL_2_WIRE_RES
•
Range
=
1e6
•
Resolution
=
1
32. Call
niDMM_Read
. Subtract the previously stored 1 M
Ω
range offset
from this measurement. Verify that the result falls between the
tolerances listed in Table 14.
33. Output 100 k
Ω
on the calibrator with no external sense or 2-wire
compensation.
34. Call
niDMM_ConfigureMeasurement
with the following
parameters:
•
Function
=
NIDMM_VAL_2_WIRE_RES
•
Range
=
100e3
•
Resolution
=
0.1
35. Call
niDMM_Read
. Subtract the previously stored 100 k
Ω
range offset
from this measurement. Verify that the result falls between the
tolerances listed in Table 14.