National Instruments Corporation
45
NI PXI-4070 Calibration Procedure
49. Call
niDMM_CalAdjustOffset
with the following parameters:
•
Mode
=
NIDMM_VAL_4_WIRE_RES
•
Range
=
1e6
•
Input Resistance
=
NIDMM_VAL_RESISTANCE_NA
50. Call
niDMM_CalAdjustOffset
with the following parameters:
•
Mode
=
NIDMM_VAL_4_WIRE_RES
•
Range
=
1e3
•
Input Resistance
=
NIDMM_VAL_RESISTANCE_NA
51. Call
niDMM_CalAdjustOffset
with the following parameters:
•
Mode
=
NIDMM_VAL_4_WIRE_RES
•
Range
=
100
•
Input Resistance
=
NIDMM_VAL_RESISTANCE_NA
52. Make the following changes to the connections between the calibrator
and the calibration cable assembly spade lugs:
a.
Remove the spade lugs labeled
HS
and
LS
from the calibrator.
Leave these spade lugs unplugged and ensure that they do not
touch each other.
b.
Reconnect the 2wr COMP+ spade lug to the SENSE HI binding
post of the calibrator.
c.
Reconnect the 2wr COMP– spade lug to the SENSE LO binding
post of the calibrator.
53. Wait two minutes for the thermal EMF to stabilize.
54. Output 0
Ω
on the calibrator with external sense and 2-wire
compensation turned on.
55. Call
niDMM_CalAdjustOffset
with the following parameters:
•
Mode
=
NIDMM_VAL_2_WIRE_RES
•
Range
=
10e6
•
Input Resistance
=
NIDMM_VAL_RESISTANCE_NA
56. Call
niDMM_CalAdjustOffset
with the following parameters:
•
Mode
=
NIDMM_VAL_2_WIRE_RES
•
Range
=
1e6
•
Input Resistance
=
NIDMM_VAL_RESISTANCE_NA
57. Call
niDMM_CalAdjustOffset
with the following parameters:
•
Mode
=
NIDMM_VAL_2_WIRE_RES
•
Range
=
100e3
•
Input Resistance
=
NIDMM_VAL_RESISTANCE_NA