Table 1.
Device Front Panel Icon Definitions
Refer to the user documentation for required maintenance measures to ensure user
safety and/or preserve the specified EMC performance.
The signal pins of this product's input/output ports can be damaged if subjected to
Electrostatic Discharge (ESD). To prevent damage, turn off power to the product
before connecting cables and employ industry-standard ESD prevention measures
during installation, maintenance, and operation.
Table 2.
NI 5624R IF Digitizer Front Panel Connectors
Label
Connector Type
Function
IF IN
SMA connector
Analog input terminal for intermediate frequency (IF)
signals.
PFI 0
SMA connector
Programmable function digital I/O (DIO) connector for
triggers or events.
CLK IN
SMA connector
Input for an external ADC Clock (Sample Clock) or
Reference Clock to the digitizer.
CLK OUT
SMA connector
Output for an ADC Clock (Sample Clock) or PLL
Reference Clock from the digitizer.
DIGITAL I/O HDMI
™
connector DIO terminal that contains 12 general-purpose DIO
signals. DIO lines are direction configurable as input or
output.
Caution
The DIGITAL I/O connector accepts a standard, third-party HDMI cable,
but the DIGITAL I/O port is not an HDMI interface. Do not connect the DIGITAL
I/O port on the NI 5624R to the HDMI port of another device. NI is not liable for
any damage resulting from such signal connections.
The following table provides information about the module front panel LEDs and the device
states they indicate.
NI PXIe-5624R Getting Started Guide
|
© National Instruments
|
9