NI PXIe-1085 Series User Manual
Shock and Vibration
Operational shock ............................................. 30 g peak, half-sine, 11 ms pulse
(Tested in accordance with IEC 60068-2-27.
Meets MIL-PRF-28800F Class 2 limits.)
Random Vibration............................................. 5 to 500 Hz, 0.3 g
rms
Acoustic Emissions
Sound Pressure Level (at Operator Position)
(Tested in accordance with ISO 7779. Meets MIL-PRF-28800F requirements.)
Auto fan (up to ~30 °C ambient) ...................... 51.2 dBA
High fan ............................................................ 64.1 dBA
Sound Power
Auto fan (up to ~30 °C ambient) ...................... 60.8 dBA
High fan ............................................................ 75.9 dBA
Caution
The protection provided by the NI PXIe-1085 Series can be impaired if it
is used in a manner not described in this document.
Note
Specifications are subject to change without notice.
Safety
This product is designed to meet the requirements of the following standards of safety for
information technology equipment:
•
IEC 61010-1, EN 61010-1
•
UL 61010-1, CSA 61010-1
Note
For UL and other safety certifications, refer to the product label or the
section.
Electromagnetic Compatibility
This product meets the requirements of the following EMC standards for electrical equipment
for measurement, control, and laboratory use:
•
EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
•
EN 55011 (CISPR 11): Group 1, Class A emissions
•
EN 55022 (CISPR 22): Class A emissions
•
EN 55024 (CISPR 24): Immunity
•
AS/NZS CISPR 11: Group 1, Class A emissions