NI 5122/5124/5142 Calibration Procedure
3.
Configure the common vertical properties using the niScope Configure Vertical VI.
4.
Configure the horizontal properties using the niScope Configure Horizontal Timing VI.
LabVIEW VI
C/C++ Function Call
Call
niScope_ConfigureChan
Characteristics
with the
following parameters:
vi
: The instrument handle from
niScope_init
channelList
:
"0"
inputImpedance
:
NISCOPE_VAL_1_MEG_OHM
maxInputFrequency
:
100,000,000
(for
NI 5122/5142),
150,000,000
(for NI 5124)
LabVIEW VI
C/C++ Function Call
Call
niScope_Configure
Vertical
with the following
parameters:
coupling
:
NISCOPE_VAL_DC
probeAttenuation
:
1.0
vi
: The instrument handle from
niScope_init
channelList
:
"0"
range
:
2.0
offset
:
0.
75
enabled
:
NISCOPE_VAL_TRUE